Table 5-40 Pulse modulation RF output at +14 dBm

Carrier frequency

RF level min.

Result (dBm)

RF level max.

(MHz)

(dBm)

 

(dBm)

 

 

 

 

 

 

 

 

32

+12.5

_______

+15.5

60

+12.5

_______

+15.5

180

+12.5

_______

+15.5

300

+12.5

_______

+15.5

420

+12.5

_______

+15.5

540

+12.5

_______

+15.5

660

+12.5

_______

+15.5

780

+12.5

_______

+15.5

900

+12.5

_______

+15.5

1020

+12.5

_______

+15.5

1140

+12.5

_______

+15.5

1200

+12.5

_______

+15.5

1201

+11.5

_______

+16.5

1260

+11.5

_______

+16.5

1380

+11.5

_______

+16.5

1500

+11.5

_______

+16.5

1620

+11.5

_______

+16.5

1740

+11.5

_______

+16.5

1860

+11.5

_______

+16.5

1980

+11.5

_______

+16.5

2220

+11.5

_______

+16.5

2340

+11.5

_______

+16.5

2400

+11.5

_______

+16.5

 

 

 

 

 

 

 

 

Table 5-41

Pulse modulation on/off ratio test

 

 

 

 

Carrier frequency

 

Pulse mod. on/off

Measured value (dB)

(MHz)

 

ratio (dB)

 

 

 

 

 

 

 

 

32

 

>45

_______

100

 

>45

_______

320

 

>45

_______

1000

 

>45

_______

1200

 

>45

_______

1500

 

>40

_______

1800

 

>40

_______

2100

 

>40

_______

2400

 

>40

_______

 

 

 

 

 

 

 

 

5-50

Page 125