(8)Set the UUT to RF level +19 dBm and repeat (2) to (5) using Table
AM scale shape
Test procedure
(1)Connect the test equipment as shown in Fig.
(2)On the UUT set:
Carr Freq | 100 MHz |
RF Level | 0 dBm |
Mod Mode | AM Internal |
Source On |
|
Mod On |
|
AM1 Level | 30% |
(3)On the modulation meter, select CAL, AM, 300 Hz 3.4 kHz LF filter.
(4)Measure the AM accuracy at the depths shown in Table
External AM frequency response (ALC off, DC coupled)
Test procedure
100 Hz to 30 kHz
(1)Connect the test equipment as shown in Fig.
(2)On the UUT set:
Carr Freq | 400 MHz |
RF Level | −4 dBm |
Mod Mode | AM External |
AM1 Level | 80% |
Source On |
|
Mod On |
|
DC coupling |
|
(3)Set the function generator to give 1 V RMS, 1 kHz sine wave.
(4)On the modulation meter select CAL, AM, 10 Hz 300 kHz filter.
(5)On the modulation meter check that the AM reading is between 76% and 84%, then set a reference using the relative function.
(6)Record the absolute reading for use in the formula in (16) below.
(7)Set the function generator to each of the frequencies shown in Table
(8)Set the UUT RF level to +7 dBm and repeat (3) to (7) using Table
(9)Set the UUT RF level to +13 dBm and repeat (3) to (7) using Table