Carrier frequency accuracy
This check provides a conventional method of checking the signal generator frequency locking circuitry. It will confirm correct operation of phase locked loops and dividers. Overall accuracy is determined by the instrument’s internal reference standard.
Specification
Range | 9 kHz to 2.4 GHz. |
Resolution | 1 Hz. |
Accuracy | Equal to the frequency standard accuracy. |
Test equipment |
|
Description
Minimum specification
Example
Frequency counter
9 kHz to 2.4 GHz
Agilent 53181A with option 030
Test procedure
Counter |
UUT |
INPUT |
RF |
OUT |
C5356 |
Fig. 5-3 Carrier frequency accuracy test set-up
(1)Connect the test equipment as shown in Fig.
(2)Connect the internal frequency standard from the UUT to the external standard input on the counter.
(3)On the UUT set:
Carr Freq | 9 kHz |
RF Level | 0 dBm |
(4)Record the frequency measured by the counter against each of the carrier frequencies shown in Table
±1 digit on the counter display.)
(5)At 1200 MHz disconnect the UUT internal frequency standard from the counter and instead apply the external reference. Check the result against the limits.
The test limits quoted are for guidance and assume that the internal frequency standard has recently been adjusted. Aging and stability have to be considered when establishing the real test limits (see 'Performance data' in Chapter 1).