VXI 3002 manual Modulation oscillator frequencies

Models: 3002

1 136
Download 136 pages 1.74 Kb
Page 96
Image 96

Modulation oscillator frequencies

Test procedure

UUT

Counter

INPUT

LF OUTPUT

C5359

Fig. 5-13 Modulation oscillator frequency test set-up

(1)Connect the test equipment as shown in Fig. 5-13.

(2)On the UUT set:

FM1 Freq

10 Hz

(3)Record the frequency measured by the counter against each of the modulation oscillator frequencies shown in Table 5-43.

Modulation oscillator distortion and LF output flatness

Test procedure

UUT

 

 

LF OUTPUT

INPUT

Audio analyzer

 

 

C2569

Fig. 5-14 Modulation oscillator distortion test set-up

(1)Connect the test equipment as shown in Fig. 5-14.

(2)On the UUT set:

FM1 Freq

1 kHz

5-21

Page 96
Image 96
VXI 3002 manual Modulation oscillator frequencies, Modulation oscillator distortion and LF output flatness