(4)Measure the level of the second and third harmonics on the spectrum analyzer at each of the carrier frequencies shown in Table
(5)Set the UUT RF level to +7 dBm and repeat (4) using Table
(6)Set the UUT RF level to +13 dBm and repeat (4) using Table
(7)Set the UUT RF level to +19 dBm and repeat (4) up to 1.2 GHz using Table
Non-harmonics
Test procedure
(1)Press CAL on the spectrum analyzer.
(2)Connect the test equipment as shown in Fig.
(3)On the UUT set:
Carr Freq | 1201 MHz |
RF Level | 0 dB |
(4)Measure the level of the
Residual FM
Test procedure
UUT |
|
RF OUT | Measuring |
RF INPUT | |
| receiver |
| C2562 |
Fig. 5-5 Residual FM test set-up
(1)Connect the test equipment as shown in Fig.
(2)On the UUT set:
Carr Freq | 1 [GHz] |
RF Level | 0 dBm |
(3)On the measuring receiver select FM, 300 Hz
(4)Measure the residual FM checking that the result is within the specification shown in Table