Teledyne 9060Z manual Filter Purging, Calibration GAS Check, Dust in the Flue GAS

Page 25

Model 9060Z Manual

high impedance alarm, ‘Probe 1 (2) High Impedance’, is not active. The actual impedance can be displayed on the lower line. It should be less than 9 KΩ at 720°C (1320°F).

Once it has been established that the impedance is normal, the offset may be set using the millivolt value marked on the oxygen probe. See chapter 6.3.1, Probe Offset. The probe offset can be tested on site. A small flow of air must be admitted to both the ‘REF’ and ‘CAL’ ports when testing the probe offset. If the probe is in the process, the air must fully purge the probe sensor without interference from the process gas sample. Teledyne Analytical Instruments probes can easily achieve this with or without a probe filter and a gas flow of only 1 to 5 SLPM (2 to 10 SCFH) for a 9060H probe and up to 20 SLPM (40 SCFH) for an unheated probe. When a stable oxygen reading has been achieved, read the ‘Probe EMF’. Enter this value into the set‐up menu #1 for the probe 1 (set‐up #2 for the second probe if it is installed).

4.17 FILTER PURGING

Purging probe filters is controlled from the ‘GAS/PURGE’ buttons on the transmitter when in ‘RUN’ mode. If ‘Automatic’ has been selected in either ‘Sol. 1 Auto/Man’ or ‘Sol. 2 Auto/Man’ in Commissioning Menus 21 or 30, pressing the GAS/PURGE button will start the automatic cycle. Pressing the button again will cancel the auto purge cycle. If ‘Automatic’ was not enabled, the solenoid will only stay open for as long as the button is pressed. Gradually adjust the purge air supply regulator, increasing the pressure until sufficient flow is obtained to clear the filter. This is best checked with a dirty filter after a period of operation, by withdrawing the probe from service and watching any build up on the filter being blown off at the set pressure. Normally 30 kPa (5 psi) is adequate but the air pressure may be set as high as 100 kPa (15 psi).

4.18 CALIBRATION GAS CHECK

If the installation has a filter purge facility, set this up first. Refer to the previous paragraph. Press the ‘GAS/PURGE 1 or ‘GAS/PURGE 2’ button while in ‘SET UP’ mode to obtain a reasonable flow through the calibration check gas flow meter. If air is being used as a calibration check gas, use the air from the regulator for filter purge. Then, when setting up a gas for calibration checking, set the pressure from the calibration gas cylinder so that it is the same as the pressure set on the air regulator. Then the setting on the rotameter / flow regulator will be the same as that for the airflow. The flow required is 1 to 5 SLPM (2 to 10 SCFH) for a 9060H probe and up to 20 SLPM (40 SCFH) for an unheated probe. Air is not the best gas for calibration checking on a zirconia sensor. The output of a zirconia sensor with air on both sides of the sensor is zero millivolts. It is better to choose a gas value which provides a reasonable output from the sensor and which is near to the process oxygen level. A cylinder with 2% oxygen in nitrogen is a commonly used calibration gas. The maximum pressure on the calibration check gas cylinder regulators is 100 kPa (15 psi).

Note: If ‘Dual probe’ was selected in commissioning menu 4, ‘Cal Gas 2’ must be connected to probe 2.

4.19 DUST IN THE FLUE GAS

For unheated probes with no filter, entrained solids or dust in the flue gas does not present a problem unless the dust, when settled, is not porous. Allow the dust in the process to build up on the probe. It will form a porous layer slowing the response time. To avoid mechanical abrasion of the electrode material in installations with unheated oxygen probes, pack ‘SAFFIL’ or equivalent alumina based ceramic fiber in the sensing holes to protect the electrode. Do not use silica based ceramic fibers such as ‘KAOWOOL’, which can attack the electrode at high temperatures. Once the dust has built up the response time of the probe will be slower.

For heated probes the preferred method of mounting for dust‐laden applications is facing vertically

25

Image 25
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices 9060Z Transmitter IntroductionSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumInstalling a 9060H Oxygen Probe Installation and CommissioningMounting the Transmitter Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Interlock Relays Connecting AN Oxygen Probe CableHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple Optional Connecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixCalibration GAS Check Filter PurgingDust in the Flue GAS Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeProbe 1 Offset Setup Mode FunctionsLower Line Items Oxygen Display Units Process AlarmsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationCalibration GAS GAS Calibration Check and PurgeFunction # Function Description Options Software Upgrades Detailed Fault Analysis TroubleshootingFirst Approach Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04