Teledyne 9060Z manual Calibration Mode, Menu Function name Range

Page 42

Model 9060Z Manual

8. CALIBRATION MODE

This chapter describes the functions available when the CALIBRATION mode is selected on the transmitter.

For specific information about calibrating the transmitter see chapter 10, Instrument Calibration.

The CALIBRATION mode is accessed by pressing and holding the SETUP key for approximately 4 seconds until the words “Calibration Menu” appears at the bottom of the display. The transmitter will return to the RUN mode when the SETUP key is pressed again or 60 seconds after the last key press.

Changing options in the CALIBRATION mode is the same as the SETUP mode. See chapter 6.2 Changing Options.

8.1 FUNCTION SUMMARY TABLE

When the transmitter is in the CALIBRATION mode the SETUP light will be on and the words “Commissioning Menu” will be at the bottom of the display. The following table shows the CALIBRATION menu functions:

Menu

Function name

Range

Default

#

 

 

value

1

Reference Voltage 1, 50mV

40.00 to 60.00mV

47.14mV

2

Reference Voltage 2, 200mV

180.00 to 210.00mV

182.24mV

3

Reference Voltage 3, 1200mV

1150.0 to 1250.0mV

1221.8mV

4

Reference Voltage 4, 2500mV

2400.0 to 2550.0mV

2489.2mV

5

Output Channel 1, Calibration

Auto Calibrated Manual Calibrated Set

Auto

 

 

4mA Set 20mA

Calibrated

6

Output Channel 1, 4mA Trim

3.00 to 5.00mA

4.00mA

7

Output Channel 1, 20mA Trim

19.00 to 21.00mA

20.00mA

8

Output Channel 2, Calibration

Auto Calibrated Manual Calibrated Set

Auto

 

 

4mA Set 20mA

Calibrated

9

Output Channel 2, 4mA Trim

3.00 to 5.00mA

4.00mA

10

Output Channel 2, 20mA Trim

19.00 to 21.00mA

20.00mA

11

Ambient Temperature Sensor Offset

‐10.0°C to 10.0°C

0.0°C

12

Low Oxygen Cal 1

80.0% to 120.0%

100.00%

13

Low Oxygen Cal 2

80.0% to 120.0%

100.00%

14

Transmitter Output Select

4‐20mA / 0‐20mA / 4‐20mA over‐

4‐20mA

 

 

scale limit / 0‐20mA over scale limit

 

15

Transmitter Output Limiting

Disabled / Hold 0mA / Hold 4mA /

Hold 0mA

 

 

Hold 20mA

 

16

Mains Voltage Detection Override

Automatic / 220/240 / 10/120

Automatic

17

Mains Frequency Detection Override

Automatic / 50Hz / 60Hz

Automatic

18

Heater SSR Selection Normal

Heater1 <‐> Heater2 Heaters <‐> Cal

Normal

 

 

Purge

 

19

SSR Fail Protection

Enabled / Disabled

Enabled

42

Image 42
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Alarm relay contacts Temperature accuracyOutput load Alarm relays PowerConnection Weight Temperature RangeHead Temperature 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting 4B Earth Connection PE Installing the Auxiliary Thermocouple4A Shield Connections Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Alarms Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkProbe or Sensor Calibration Onnecting PowerCommissioning ‐ RUN Mode For more details see AppendixFilter Purging Calibration GAS CheckDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayOxygen Display Units Keypad in RUN modeKey text RUN mode White text Setup mode Black text KeypadKeypad in Setup / Commissioning / Calibration modes Enter key Function up and Function down keysOption up and Option down keys Probe impedance keyChanging Options Setup ModeFunction Summary Table Function name RangeSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Calibration Freezes Outputs Temperature UnitsFlue Pressure Units and Value Solenoid 1 & 2 OperationAlarm Relay 1, 2 and 3 Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Common Alarm Relay FunctionReference Air Pump Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Cleaning MaintenanceTransmitter Maintenance Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the N/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the S/C atom ratio in the fuelProbe EMF Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 26.2 13.120.7 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x03 0x41 Special Instruction Function0x02 0x04