Teledyne 9060Z manual Setup Mode Functions, Probe 1 Offset, Lower Line Items

Page 32

Model 9060Z Manual

6.3 SETUP MODE FUNCTIONS

6.3.1 Probe 1 Offset

Each Teledyne Analytical Instruments probe has an offset calibration value printed on a tag that is attached to the probe when it is dispatched. To achieve accurate measurement the offset value must be entered with the same polarity as it is printed on this label. The offset value should be within ±1.0mV

NOTE: An offset of 1.0mV will change the oxygen reading by approximately 1% oxygen when the probe is in ambient air. However, as the process oxygen measurement drops, this offset will have a reduced affect. At a process gas oxygen concentration of 2%, the 1.0mV offset error will only change the reading by 0.1% oxygen.

If in any doubt about the correct offset value, set it to 0.0mV.

The function ’Probe 2 Offset’ will only appear if the transmitter has been configured for 2 probes. (see chapter 7.2.3 Number of Probes )

6.3.2 Lower Line Items

This function allows the operator to change the items that are available to be displayed on the lower line of the transmitter in RUN mode. If the word “Enabled” appears on the display for a selected lower line option, the measurement will be available to be shown on the display in the RUN mode by scrolling through the list using the DISPLAY UP and DISPLAY DOWN keys.

A lower line measurement can be “Enabled” or disabled by pressing the ENTER key.

32

Image 32
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Temperature accuracy Output load Alarm relaysAlarm relay contacts PowerTemperature Range Head TemperatureConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting Installing the Auxiliary Thermocouple 4A Shield Connections4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Auxiliary Thermocouple Optional Connecting the Output ChannelsConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkOnnecting Power Commissioning ‐ RUN ModeProbe or Sensor Calibration For more details see AppendixDust in the Flue GAS Filter PurgingCalibration GAS Check Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayKeypad in RUN mode Key text RUN mode White text Setup mode Black textOxygen Display Units Keypad Keypad in Setup / Commissioning / Calibration modes Function up and Function down keys Option up and Option down keysEnter key Probe impedance keySetup Mode Function Summary TableChanging Options Function name RangeLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Temperature Units Flue Pressure Units and ValueCalibration Freezes Outputs Solenoid 1 & 2 OperationOxygen Content Calibration Gas 1 Maximum Calibration Gas 1 & 2 Positive / Negative ErrorAlarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionOperation of the Alarm Relays when an Alarm is Accepted Selecting the Correct FuelReference Air Pump Options Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Maintenance Transmitter MaintenanceCleaning Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the H/C atom ratio in the fuel Is the O/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelAppendix 2, Probe EMF Tables Zirconia Oxygen Sensor Output mV Probe Type 9060H, SensorProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 13.1 20.726.2 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x41 Special Instruction Function 0x020x03 0x04