Teledyne 9060Z manual Appendix 2, Probe EMF Tables, Probe EMF @ 720C 1320F

Page 58

Model 9060Z Manual

APPENDIX 2, PROBE EMF TABLES

ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060H, SENSOR

 

 

OXYGEN

 

 

 

 

Probe EMF

 

 

OXYGEN

 

 

Probe EMF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

%

 

 

 

@ 720°C (1320°F)

 

 

%

 

 

@ 720°C (1320°F)

 

 

 

20.95

 

 

0

 

 

10

 

 

15.93

 

 

 

 

 

 

 

 

 

20.5

 

 

0.517

 

 

9.5

 

17.033

 

 

 

20

 

 

1.046

 

 

9

 

 

18.196

 

 

19.5

 

 

1.589

 

 

8.5

 

19.426

 

 

 

19

 

 

2.147

 

 

8

 

 

20.73

 

 

18.5

 

 

2.719

 

 

7.5

 

22.12

 

 

 

18

 

 

3.306

 

 

7

 

 

23.607

 

 

17.5

 

 

3.911

 

 

6.5

 

25.204

 

 

 

17

 

 

4.533

 

 

6

 

 

26.93

 

 

16.5

 

 

5.173

 

 

5.5

 

28.808

 

 

 

16

 

 

5.834

 

 

5

 

 

30.867

 

 

15.5

 

 

6.515

 

 

4.5

 

33.145

 

 

 

15

 

 

7.219

 

 

4

 

 

35.695

 

 

14.5

 

 

7.947

 

 

3.5

 

38.59

 

 

 

14

 

 

8.7

 

 

3

 

 

41.94

 

 

13.5

 

 

9.481

 

 

2.5

 

45.913

 

 

 

13

 

 

10.292

 

 

2

 

 

50.797

 

 

12.5

 

 

11.134

 

 

1.5

 

57.135

 

 

 

12

 

 

12.011

 

 

0.9

 

 

66.182

 

 

11.5

 

 

12.925

 

 

0.4

 

82.168

 

 

 

11

 

 

13.881

 

 

0.2

 

 

99.518

 

 

10.5

 

 

14.881

 

 

0.1

 

114.347

 

 

'K' Type TC 29.965 mV @ 720°C (1320°F)

 

 

 

 

 

These tables are based on the Nernst equation: O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

EMF =

 

 

 

 

 

.

 

 

Where T is temperature° K

 

 

 

.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

58

Image 58
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices 9060Z Transmitter IntroductionSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Alarm relay contacts Temperature accuracyOutput load Alarm relays PowerConnection Weight Temperature RangeHead Temperature 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstalling a 9060H Oxygen Probe Installation and CommissioningMounting the Transmitter Oxygen Probe Mounting 4B Earth Connection PE Installing the Auxiliary Thermocouple4A Shield Connections Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Interlock Relays Connecting AN Oxygen Probe CableHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Alarms Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkProbe or Sensor Calibration Onnecting PowerCommissioning ‐ RUN Mode For more details see AppendixCalibration GAS Check Filter PurgingDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayOxygen Display Units Keypad in RUN modeKey text RUN mode White text Setup mode Black text KeypadKeypad in Setup / Commissioning / Calibration modes Enter key Function up and Function down keysOption up and Option down keys Probe impedance keyChanging Options Setup ModeFunction Summary Table Function name RangeProbe 1 Offset Setup Mode FunctionsLower Line Items Oxygen Display Units Process AlarmsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Calibration Freezes Outputs Temperature UnitsFlue Pressure Units and Value Solenoid 1 & 2 OperationAlarm Relay 1, 2 and 3 Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Common Alarm Relay FunctionReference Air Pump Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartCalibration GAS GAS Calibration Check and PurgeFunction # Function Description Options Software Upgrades Detailed Fault Analysis TroubleshootingFirst Approach Cleaning MaintenanceTransmitter Maintenance Replacement Parts Index Auxiliary ThermocoupleTroubleshooting Is the N/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the S/C atom ratio in the fuelProbe EMF Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 26.2 13.120.7 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x03 0x41 Special Instruction Function0x02 0x04