Teledyne 9060Z manual Alarm Relay Options

Page 48

Model 9060Z Manual

level. With heated probes this relay will be de‐energised while the probe is heating up from ambient, making the contacts open circuit.

30.‘Cal 1 in Progress’

31.‘Cal 2 in Progress’

A calibration check is occurring, either manual or automatic mode.

32.‘Purge 1 in Progress’

33.‘Purge 2 in Progress’

A probe purge is occurring, either manual or automatic mode.

9.3 ALARM RELAY OPTIONS

The three process alarm relays (relays 1 to 3) are user defined. The relay action will depend on the selections made in the COMMISSIONING menu functions #45 to #48.

The relay contacts are all designed to be “Fail Safe”. That is, they are

open when the transmitter power is off

closed when there is NOT an alarm condition

open when an alarm occurs

The contact will close circuit again (relay energised) when the alarm condition is acknowledged.

Choose to have the relays react to any or all of the following alarms or warnings –

Oxygen on probe 1 Low

Alarm

Oxygen on probe 2 Low

Alarm

Oxygen on probe 1 Very Low

Alarm

Oxygen on probe 2 Very Low

Alarm

Oxygen on probe 1 High

Alarm

Oxygen on probe 2 High

Alarm

Oxygen Deviation High

Alarm

Probe 1 Temperature Low

Warning

Probe 2 Temperature Low

Warning

Calibration check on probe 1 in

Warning

Progress

 

Calibration check on probe 2 in

Warning

Progress

 

Purge on probe 1 in Progress

Warning

Purge on probe 2 in Progress

Warning

In addition to the above process alarms that can activate the process alarm relays, any of the common alarm relay events that have been taken off the common alarm list in COMMISSIONING menu function #49 will appear on the process alarm relay lists in function #46 to #48.

48

Image 48
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Temperature accuracy Output load Alarm relaysAlarm relay contacts PowerTemperature Range Head TemperatureConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting Installing the Auxiliary Thermocouple 4A Shield Connections4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Auxiliary Thermocouple Optional Connecting the Output ChannelsConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkOnnecting Power Commissioning ‐ RUN ModeProbe or Sensor Calibration For more details see AppendixFilter Purging Calibration GAS CheckDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayKeypad in RUN mode Key text RUN mode White text Setup mode Black textOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Function up and Function down keys Option up and Option down keysEnter key Probe impedance keySetup Mode Function Summary TableChanging Options Function name RangeSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Temperature Units Flue Pressure Units and ValueCalibration Freezes Outputs Solenoid 1 & 2 OperationOxygen Content Calibration Gas 1 Maximum Calibration Gas 1 & 2 Positive / Negative ErrorAlarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionOperation of the Alarm Relays when an Alarm is Accepted Selecting the Correct FuelReference Air Pump Options Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Maintenance Transmitter MaintenanceCleaning Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the H/C atom ratio in the fuel Is the O/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelAppendix 2, Probe EMF Tables Zirconia Oxygen Sensor Output mV Probe Type 9060H, SensorProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 13.1 20.726.2 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x41 Special Instruction Function 0x020x03 0x04