13. TROUBLESHOOTING

The Teledyne Analytical Instruments oxygen transmitter has proved an extremely reliable instrument. The trouble shooting guide here is based on an analysis of the potential problem that may occur after many years of operation in the field. A current list of problems and solutions can be found on the Teledyne Analytical Instruments web site at www.teledyne‐ai.com when any are identified.

13.1 FIRST APPROACH

The probe will not heat up.

Turn the power off, remove and check 2 x 20mm glass fuses. If either of them is blown, check the probe heater resistance between the two white wires that come from the probe. It should be 110 +/‐ 15 ohms, and both wires should be open circuit to earth. If not, replace the probe.

Is the ‘B’ shown on the bottom left hand corner of the display? If not, check that the burner is enabled with either a link between terminals 10&11, BURNER INPUT, or is enabled with a connection to a voltage‐free contact from the main fuel valve to these terminals.

The display is blank and there is no backlight on.

Measure the power supply voltages at the test points labeled COMM and +5v at the right hand side of the main PCB 1732‐1. It should be 5 +/‐ 0.15v.

Turn off the power and remove the main shield. Check that the plug‐in fuse FS1 has not blown. It should only be replaced with a 1A, 250v fuse if it has blown.

Is the mains power being supplied to terminals 36&37 and is it between 100 and 240VAC? If may be necessary to replace the switch mode power supply, PS5, APC‐5S.

An ‘ADC Calibration Fail’ alarm has occurred.

Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the voltages on the ‘Acom’ test points. With the negative lead on the centre terminal the other 2 terminals should measure 12 +/‐0.3v. If not, replace the small DC‐DC converter PS4.

An ‘Output 1 (2) Failure’ alarm has occurred.

Turn the transmitter off and remove the main shield. Turn the transmitter back on and measure the voltages on the ‘D1com’ test points. With the negative lead on the centre terminal the other 2 terminals should measure 12 +/‐0.3v. If not, replace the small DC‐DC converter PS4.

A ‘Heater 1 (2) Failure’ alarm has occurred.

The software has found that the SSR4 has failed. If only one probe is being used and you need to gets the transmitter back working quickly, use the CALIBRATION menu function 18, Heater SSR Select, to change the heater output terminals. See chapter 8.2.8, Heater SSR fault correction.

A ‘BBRAM Fail’ alarm has occurred

Replace the BBRAM, MEM1 on the 1732‐1 main PCB.

13.2 DETAILED FAULT ANALYSIS

The 9060Z transmitter has a diagnostics mode built into the software. This mode allows detailed analysis of the hardware of the transmitter, but does require a level of competence in electronics. The diagnostics mode is selectable by turning the ‘Diagnostics’ DIP switch to ON and then turn on the power. There is a separate Diagnostics Mode Manual available that describes its use. Ask Teledyne Analytical Instruments for more details.

53

Image 53
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple OptionalConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04