Model 9060Z Manual

The on‐line auto purge and calibration check system is optional. For details on its operation see chapters 4.24, 4.25 and chapter 11.

To automatically sense a blocked probe filter, a flow switch should be installed in the ‘purge’ line to the probe ‘CAL’ port. It should be adjusted so that it energizes just below the purge flow with new or clean filters installed. The flow switch contacts should be connected to terminals 16 & 17 (FLOW SW). If the filter is still blocked or partly blocked after an auto purge cycle, the flow switch will not energize and will therefore cause a ‘Probe1 (2) Filter Blocked’ alarm.

After installation the purge/cal system should be tested thoroughly for leaks. Any leaks can cause significant errors if the flue is at negative pressure. If the flue is at positive pressure, an outward leak can cause corrosion in the purge/cal system piping and fittings.

If probe/filter purging is required but a “Probe1 (2) Filter Blocked” alarm is not required, do not install the flow switch but link terminals 16 &17.

Automatic Purge &

Calibration check

System Wiring

Schematic

Automatic Purge &

Calibration check

System Piping

Schematic

4.12 CONNECTING REFERENCE AIR

For oxygen probes, a 1/4” tube connector on the transmitter should be connected via a nylon, copper or stainless steel tube to the ‘REF’ connector on the probe. If two probes are being used, a “T” fitting must be supplied to provide reference air supply to both probes.

4.13 CONNECTING THE TRANSMITTER TO A MODBUS™ NETWORK

The transmitter can be networked to other transmitters and to a network master. The network uses the transmitter’s RS485 port. Up to 31 transmitters can be connected to the network, and can be interrogated by the Network Master.

NOTE: Hardware Protocol Selection The communications port of the 9060Z can be configured to communicate on either RS232 or RS485. If the transmitter is to be used on a MODBUS network, the 9060Z transmitter must be set to RS485. For further details see chapter 7.2.25, Communication Port Options.

NOTE: Terminating Resistor There is a terminating 100 ohm resistor fitted to the 1732‐1 PCB. Link LK3 (TERM), in the bottom left‐hand corner of the PCB, is used to connect the terminating resistor. Link LK3 must be removed on all transmitters except the transmitter on the end of the network line. If the

23

Image 23
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Power Temperature accuracyOutput load Alarm relays Alarm relay contacts9060H 9060UL/UH Application Temperature RangeHead Temperature Connection WeightHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting Electrical Connections Installing the Auxiliary Thermocouple4A Shield Connections 4B Earth Connection PEModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Automatic Purge and Calibration Check System Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the AlarmsConnecting the Transmitter to a Modbus Network Connecting Reference AIRFor more details see Appendix Onnecting PowerCommissioning ‐ RUN Mode Probe or Sensor CalibrationDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKeypad Keypad in RUN modeKey text RUN mode White text Setup mode Black text Oxygen Display UnitsKeypad in Setup / Commissioning / Calibration modes Probe impedance key Function up and Function down keysOption up and Option down keys Enter keyFunction name Range Setup ModeFunction Summary Table Changing OptionsLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Solenoid 1 & 2 Operation Temperature UnitsFlue Pressure Units and Value Calibration Freezes OutputsCommon Alarm Relay Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Alarm Relay 1, 2 and 3 FunctionFunction Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Reference Air Pump OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Replacement Parts MaintenanceTransmitter Maintenance CleaningAuxiliary Thermocouple IndexTroubleshooting Is the S/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the N/C atom ratio in the fuelProbe EMF @ 720C 1320F Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMFZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 33.8 13.120.7 26.2Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x04 0x41 Special Instruction Function0x02 0x03