Teledyne 9060Z manual Oxygen Display Units, Keypad in RUN mode

Page 28

Model 9060Z Manual

5.2 OXYGEN DISPLAY UNITS

The measured oxygen content displayed on the top line of the transmitter for probe 1 and on the lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen Display Units for how to change the oxygen display.

The format of the oxygen display changes to maintain the best resolution for the measurement. If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format:

Range

Display format

30.0 to 100.0%

##.#%

1.00 to 29.99%

##.##%

0.100 to 0.999%

#.###%

< 0.100%

scientific notation (#.## x 10 ‐ ## % )

If ‘Oxygen ppm’ is selected the transmitter will show the oxygen in ppm format between 10 and 9,999ppm. Outside of this range the oxygen will be shown in % format described above.

Display format for ‘Oxygen ppm’:

Range

Display format

> 10,000ppm (1%)

% format (see ‘Oxygen %’ section above)

1,000 to 9,999ppm

#,### ppm

100.0 to 999.9ppm

###.# ppm

10.00 to 99.99ppm

##.## ppm

< 10ppm

scientific notation (#.## x 10 ‐ ## % )

5.3 KEYPAD

There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or SETUP mode.

Key text

RUN mode (WHITE text)

SETUP mode (BLACK text)

SETUP / RUN

Enter SETUP mode

Return to RUN mode

DISPLAY / FUNCTION UP

Display Next Lower Line Item

Next Function

DISPLAY / FUNCTION

Display Prev. Lower Line Item

Previous Function

DOWN

 

 

ALARM / OPTION UP

Next Alarm

Next / Increment Option

ALARM / OPTION DOWN

Previous Alarm

Previous / Decrement Option

ALARM ACCEPT / ENTER

Acknowledge Displayed Alarm

Save Current Option

GAS 1 PURGE 1 / SENS IMP

Gas 1 / Purge 1 manual activate

Check Probe impedance

GAS 2 PURGE 2 / AUTO CAL

Gas 2 / Purge 2 manual activate

Manually Activate Output Calibration

5.3.1 Keypad in RUN mode

When the transmitter is turned on and has completed its start‐up procedure, it will enter RUN mode and enable the keypad. The key functions in this mode are as follows;

28

Image 28
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices 9060Z Transmitter IntroductionSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Temperature accuracy Output load Alarm relaysAlarm relay contacts PowerTemperature Range Head TemperatureConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstalling a 9060H Oxygen Probe Installation and CommissioningMounting the Transmitter Oxygen Probe Mounting Installing the Auxiliary Thermocouple 4A Shield Connections4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Interlock Relays Connecting AN Oxygen Probe CableHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Auxiliary Thermocouple Optional Connecting the Output ChannelsConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkOnnecting Power Commissioning ‐ RUN ModeProbe or Sensor Calibration For more details see AppendixCalibration GAS Check Filter PurgingDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayKeypad in RUN mode Key text RUN mode White text Setup mode Black textOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Function up and Function down keys Option up and Option down keysEnter key Probe impedance keySetup Mode Function Summary TableChanging Options Function name RangeProbe 1 Offset Setup Mode FunctionsLower Line Items Oxygen Display Units Process AlarmsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Temperature Units Flue Pressure Units and ValueCalibration Freezes Outputs Solenoid 1 & 2 OperationOxygen Content Calibration Gas 1 Maximum Calibration Gas 1 & 2 Positive / Negative ErrorAlarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionOperation of the Alarm Relays when an Alarm is Accepted Selecting the Correct FuelReference Air Pump Options Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartCalibration GAS GAS Calibration Check and PurgeFunction # Function Description Options Software Upgrades Detailed Fault Analysis TroubleshootingFirst Approach Maintenance Transmitter MaintenanceCleaning Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the H/C atom ratio in the fuel Is the O/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelAppendix 2, Probe EMF Tables Zirconia Oxygen Sensor Output mV Probe Type 9060H, SensorProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 13.1 20.726.2 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x41 Special Instruction Function 0x020x03 0x04