Model 9060Z Manual
5.2 OXYGEN DISPLAY UNITS
The measured oxygen content displayed on the top line of the transmitter for probe 1 and on the lower line for probe 2 can be shown in either % or parts per million (ppm). See chapter 6.3.3 Oxygen Display Units for how to change the oxygen display.
The format of the oxygen display changes to maintain the best resolution for the measurement. If ‘Oxygen %’ is selected the transmitter will display oxygen in the following format:
Range | Display format |
30.0 to 100.0% | ##.#% |
1.00 to 29.99% | ##.##% |
0.100 to 0.999% | #.###% |
< 0.100% | scientific notation (#.## x 10 ‐ ## % ) |
If ‘Oxygen ppm’ is selected the transmitter will show the oxygen in ppm format between 10 and 9,999ppm. Outside of this range the oxygen will be shown in % format described above.
Display format for ‘Oxygen ppm’:
Range | Display format |
> 10,000ppm (1%) | % format (see ‘Oxygen %’ section above) |
1,000 to 9,999ppm | #,### ppm |
100.0 to 999.9ppm | ###.# ppm |
10.00 to 99.99ppm | ##.## ppm |
< 10ppm | scientific notation (#.## x 10 ‐ ## % ) |
5.3 KEYPAD
There are eight keys built into the decal on the outside of the door of the 9060Z transmitter. The key functions are written in BLACK and WHITE to identify the function of the key in either RUN mode or SETUP mode.
Key text | RUN mode (WHITE text) | SETUP mode (BLACK text) |
SETUP / RUN | Enter SETUP mode | Return to RUN mode |
DISPLAY / FUNCTION UP | Display Next Lower Line Item | Next Function |
DISPLAY / FUNCTION | Display Prev. Lower Line Item | Previous Function |
DOWN |
|
|
ALARM / OPTION UP | Next Alarm | Next / Increment Option |
ALARM / OPTION DOWN | Previous Alarm | Previous / Decrement Option |
ALARM ACCEPT / ENTER | Acknowledge Displayed Alarm | Save Current Option |
GAS 1 PURGE 1 / SENS IMP | Gas 1 / Purge 1 manual activate | Check Probe impedance |
GAS 2 PURGE 2 / AUTO CAL | Gas 2 / Purge 2 manual activate | Manually Activate Output Calibration |
5.3.1 Keypad in RUN mode
When the transmitter is turned on and has completed its start‐up procedure, it will enter RUN mode and enable the keypad. The key functions in this mode are as follows;