Teledyne 9060Z manual Keypad in Setup / Commissioning / Calibration modes

Page 29

Model 9060Z Manual

SETUP / RUN key

Pressing this key will put the transmitter into one of the menu‐driven SETUP, COMMISSIONING or CALIBRATION modes (see chapters 6, 7 and 8 for details). The function of all the keys will then change to the functions that they have in the SETUP mode.

DISPLAY UP / DOWN keys

The display keys are used to scroll the lower line up and down through the variety of measurements that are available on the lower line. For a complete list of options see chapter 6.3.2, Lower Line Items.

ALARM UP key

If there is either a new alarm or an active alarm the ALARM UP key can be pressed to examine the alarm status. The alarm light will be flashing if there is a new alarm or steady if there is an existing alarm. For more details on the alarm mode and keys see chapter 9, Alarms. The SETUP light will flash slowly to show that the transmitter is now in the alarm display mode.

ALARM DOWN key

When the transmitter is displaying active alarms (the ALARM UP key has been pressed), the ALARM DOWN key and the ALARM UP key allow the operator to examine the date / time of the alarm and the date / time that the alarm was acknowledged.

If the transmitter was in RUN mode when this key is pressed it will go into the alarm log display mode. See chapter 9 for more details.

ALARM ACCEPT key

The ALARM ACCEPT key is used to accept a new alarm (see chapter 9, Alarms).

GAS 1 / PURGE 1 and GAS 2 / PURGE 2 keys

These two keys are used to turn on the gas / purge solenoids. When the transmitter is in the manual cal / purge mode (Commissioning functions #21 and 30) the solenoid will be activated for as long as the key is pressed. When the transmitter is in the auto cal / purge mode the automatic cal / purge cycle is started. The cycle can be stopped by pressing the same key again.

5.3.2 Keypad in SETUP / COMMISSIONING / CALIBRATION modes

NOTE: In order to enter either COMMISSIONING or CALIBRATION modes, the corresponding DIP switch on the 1732‐1 PCB must be turned on.

From RUN mode, if the SETUP / RUN key is pressed once, the transmitter will go into the SETUP mode.

From RUN mode, if the SETUP / RUN key is pressed and held for 2 seconds, the transmitter will go into the COMMISSIONING mode.

From RUN mode, if the SETUP / RUN key is pressed and held for 4 seconds, the transmitter will go into the CALIBRATION mode.

The following key functions are then available in all of the above modes.

SETUP / RUN key

Pressing the SETUP / RUN key while in any of the set up modes will return the transmitter to the RUN mode.

29

Image 29
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple OptionalConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04