Teledyne 9060Z manual Installing the Auxiliary Thermocouple, 4A Shield Connections

Page 17

Model 9060Z Manual

4.3 INSTALLING THE AUXILIARY THERMOCOUPLE

Weld a 1/2 inch BSP mounting socket to the flue within about 300 mm (12”), and upstream of the oxygen probe. The thermocouple should be of similar length to the oxygen probe to prevent flue temperature distribution errors.

The thermocouple should be connected to terminals 7 & 8. These terminals will not be available for an auxiliary thermocouple if a second probe has been installed.

4.4A SHIELD CONNECTIONS

All external wiring to the 9060Z transmitter should be shielded. Do not connect shields at the field end. Simply clip off and insulate. There are two M4 earth screw terminals available in the 9060Z transmitter. An extra terminal strip may be required to connect all shields together. This should be supplied by the installer.

4.4B EARTH CONNECTION (PE)

The PE (protective earth) input connection must be made to the earth stud on the right hand side of cabinet. The PE input connection must be the first connection onto the earth stud and it must be secured by a separate nut and spring washer. All other earth connections (bonding connections) can be made on either of the two earth studs in the base of the cabinet.

The transmitter MUST be securely earthed.

Assembly of the PE and bonding connections on the earth stud

4.5 ELECTRICAL CONNECTIONS

All wiring should comply with local electrical codes. The printed circuit boards are fully floating above earth. The incoming safety earth (PE) must be connected to the primary earth stud in the right hand side of thebase cabinet.

The local fuse for the mains power supply, the isolation switch and the supply wiring must all comply with the electrical safety codes and must only be installed by qualified technicians. All earth and shield connections should be connected to the earth screws inside the case.

17

Image 17
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple OptionalConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04