Teledyne manual Model 9060Z Manual

Page 46

Model 9060Z Manual

4. ‘Probe 2 High Impedance’

Oxygen probe or electrode failure. The sensor has a high impedance measurement. Replace the probe. This alarm is inhibited when the probes are less than 650°C (1200°F).

5.‘Probe 1 TC Open Circuit’

6.‘Probe 2 TC Open Circuit’

Probe thermocouple is open circuit. The heater in heated probes will switch off.

7. ‘Aux TC Open Circuit’

The AUX thermocouple is open circuit. If the thermocouple is not needed, select “Disabled” in COMMISSIONING menu function #9, “Aux TC Type” or place a short circuit between terminals 7 & 8.

8. ‘Ref Air Pump Fail’

The reference air pump in the transmitter has been either unplugged or has gone open circuit (<20mA). Replace the reference air pump.

9. ‘Ref Air Pump Overload

The reference air pump in the transmitter has stalled and is drawing in excess of 300mA. The transmitter will attempt to restart the pump every minute for 1 second. Replace the reference air pump.

10. ‘ADC Calibration Fail’

The analogue to digital converter has been found to fall outside the normal calibration specifications. In this case the probe heater will automatically be turned off.

11. ‘Alarm Log Fail’

The alarm history is kept in an EEROM. This alarm will be raised if this memory device fails.

12.‘Output 1 Failure’

13.‘Output 2 Failure’

The digital to analogue and voltage isolator circuit has been found to fall outside the normal calibration specifications. This check is only performed when the ‘AUTO CAL’ button is pressed. See chapter 10.1.2.

14.‘Heater 1 SSR Failure’

15.‘Heater 2 SSR Failure’

One of the heater power control devices (SSR) has been found to have failed. See chapter 8.2.8, Heater SSR fault correction.

16. ‘Heater SSR Leakage’

One of the heater power control devices (SSR) has been found to have failed but the transmitter cannot determine which one(s) has failed.

17.‘Probe 1 Filter Blocked’

18.‘Probe 2 Filter Blocked’

Blocked probe filter. This test is only performed when automatic purging of the probe is selected. See chapter 11. This alarm will not reset until the next purge cycle. The cycle can be initiated manually or automatically.

46

Image 46
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices 9060Z Transmitter IntroductionSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Alarm relay contacts Temperature accuracyOutput load Alarm relays PowerConnection Weight Temperature RangeHead Temperature 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstalling a 9060H Oxygen Probe Installation and CommissioningMounting the Transmitter Oxygen Probe Mounting 4B Earth Connection PE Installing the Auxiliary Thermocouple4A Shield Connections Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Interlock Relays Connecting AN Oxygen Probe CableHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Alarms Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkProbe or Sensor Calibration Onnecting PowerCommissioning ‐ RUN Mode For more details see AppendixCalibration GAS Check Filter PurgingDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayOxygen Display Units Keypad in RUN modeKey text RUN mode White text Setup mode Black text KeypadKeypad in Setup / Commissioning / Calibration modes Enter key Function up and Function down keysOption up and Option down keys Probe impedance keyChanging Options Setup ModeFunction Summary Table Function name RangeProbe 1 Offset Setup Mode FunctionsLower Line Items Oxygen Display Units Process AlarmsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Calibration Freezes Outputs Temperature UnitsFlue Pressure Units and Value Solenoid 1 & 2 OperationAlarm Relay 1, 2 and 3 Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Common Alarm Relay FunctionReference Air Pump Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartCalibration GAS GAS Calibration Check and PurgeFunction # Function Description Options Software Upgrades Detailed Fault Analysis TroubleshootingFirst Approach Cleaning MaintenanceTransmitter Maintenance Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the N/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the S/C atom ratio in the fuelProbe EMF Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 26.2 13.120.7 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x03 0x41 Special Instruction Function0x02 0x04