Teledyne 9060Z manual GAS Calibration Check and Purge, Calibration GAS

Page 51

Model 9060Z Manual

11. GAS CALIBRATION CHECK AND PURGE

The oxygen sensor that is used in the Teledyne Analytical Instruments oxygen probe is extremely predictable, stable and reliable. For this reason, the calibration of a Teledyne Analytical Instruments oxygen system does not require the use of calibration gases. However, all Teledyne Analytical Instruments oxygen probes have a built in gas connection that does allow the accuracy of the probe to be checked in‐situ. This chapter describes the operation of this gas checking system.

The 9060Z has a timer and solenoid driving system that can be configured to admit a certified calibration gas into the probe or an air supply to purge the probe filters through the gas connection. Both the calibration gas and the filter purge gas must be piped to the port on the probe labeled “CAL”. There are two solenoids drivers in the 9060Z transmitter. They can be used for any combinations of gas checking and probe filter purging. The only limitation is that if the transmitter is in dual probe mode, then solenoid #1 must be wired and plumbed to the probe #1 and solenoid #2 must be wired and plumbed to probe #2.

The transmitter can also be configured to be in a MANUAL or AUTOMATIC purge and gas check mode.

Function #

Function Description

Options

 

 

 

20 (29)

Solenoid 1

(2)

Operation

Calibration Gas / Purge

21 (30)

Solenoid 1

(2) Auto / Manual

Automatic / Manual

22 (31)

Solenoid 1

(2)

Start Time

00:00 to 23:45

 

23 (32)

Solenoid 1

(2)

Period

1 hour – 7 days

 

24 (33)

Solenoid 1 (2) Duration

1 – 90 seconds

 

25 (34)

Solenoid 1 (2) Post Freeze

5 – 300 seconds

 

26 (35)

Oxygen Content Cal Gas 1 (2)

0.1 – 20.9%

 

 

27 (36)

Max Cal Gas 1 (2) Positive Error

0.1% to 3.0%

 

 

28 (37)

Max Cal Gas 1 (2) Negative Error

0.1% to 3.0%

 

 

Functions available if either Calibration Gas or Purge and Manual are selected

11.1 PURGE

Functions available if Purge and Automatic are selected

Functions available if Calibration Gas and Automatic are selected

For processes that have particulate present in the gas stream, probe filters are necessary and should be back‐purged with sufficient frequency to avoid blocked filters. Select ‘Purge’ in function #20 (29) and ‘Automatic’ in function #30 and then set the times in function #22 (31) to #25 (34) The outputs can be frozen during purging if selected in COMMISSIONING menu function #19. A flow switch can be wired to terminals 16 and 17. If the flow is not sufficient at the end of the purge cycle to close the contacts of the flow switch a ‘Probe 1 (2) Filter Blocked’ alarm is generated. If a flow switch is not being used a short circuit should be placed across terminals 16 and 17 to suppress the ‘Probe 1 (2) Filter Blocked’ alarm.

11.2 CALIBRATION GAS

If it is required to automatically check the oxygen probe and transmitter on line, select ‘Calibration Gas’ in function #20 (29) and ‘Automatic’ in function #30 and then set the times in function #22 (31) to #25 (34) and the gas levels in function #26 (35) to 28 (37). If the transmitter measurement does not fall within the limits set in function #26 (35) to 28 (37) a ‘Gas1 (2) Calibration Error’ alarm will be generated. If the transmitter has ‘Automatic’ selected in function #21 (30) and one of the GAS/PURGE keys is pressed in RUN mode, the automatic cycle will be started and the alarms will be checked. The cycle can be stopped at any time by pressing the same key during the cycle.

51

Image 51
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Power Temperature accuracyOutput load Alarm relays Alarm relay contacts9060H 9060UL/UH Application Temperature RangeHead Temperature Connection WeightHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting Electrical Connections Installing the Auxiliary Thermocouple4A Shield Connections 4B Earth Connection PEModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Automatic Purge and Calibration Check System Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the AlarmsConnecting the Transmitter to a Modbus Network Connecting Reference AIRFor more details see Appendix Onnecting PowerCommissioning ‐ RUN Mode Probe or Sensor CalibrationFilter Purging Calibration GAS CheckDust in the Flue GAS Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKeypad Keypad in RUN modeKey text RUN mode White text Setup mode Black text Oxygen Display UnitsKeypad in Setup / Commissioning / Calibration modes Probe impedance key Function up and Function down keysOption up and Option down keys Enter keyFunction name Range Setup ModeFunction Summary Table Changing OptionsSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Solenoid 1 & 2 Operation Temperature UnitsFlue Pressure Units and Value Calibration Freezes OutputsCommon Alarm Relay Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Alarm Relay 1, 2 and 3 FunctionFunction Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Reference Air Pump OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Replacement Parts MaintenanceTransmitter Maintenance CleaningAuxiliary Thermocouple IndexTroubleshooting Is the S/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the N/C atom ratio in the fuelProbe EMF @ 720C 1320F Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMFZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 33.8 13.120.7 26.2Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x04 0x41 Special Instruction Function0x02 0x03