APPENDIX 1, CONSTITUENT VALUES FOR VARIOUS FUELS

If the transmitter is set up to have readout or output of combustibles, oxygen deficiency, maximum carbon dioxide or combustion efficiency, then the fuel constituents must be entered. Any or all of the variables can be modified and entered in COMMISSIONING menu functions #50 to 56. (See chapter 7.2.23). Your fuel supplier or chemist should be able to give you all these values.

A

is the heat of combustion of the fuel per gram atom of contained carbon.

H

is the H/C atom ratio in the fuel.

O

is the O/C atom ratio in the fuel.

N

is the N/C atom ratio in the fuel.

S

is the S/C atom ratio in the fuel.

M

is the ratio of H2O molecules to C atoms in the fuel

FUEL

Blast Furnace Gas

Coke Oven Gas

Producer Gas

Natural Gas

Propane, Natural

Butane, Refinery

Methanol

Gasoline, Motor

No 1 Distillate Oil

No 2 Distillate Oil

No 4 Fuel Oil

No 5 Residual Oil

No 6 Residual Oil

Wood, Non‐Resinous

Coal, Bituminous

Coal, Anthracite

Coke

A

H

O

N

S

M

50.55

0.08

 

1.3

3.08

b

a

256.88

5.6

 

0.25

0.23

b

a

101.98

1.18

 

1.02

2.9

b

a

209.9

3.86

 

0

0.1

0

0

176.4

2.69

 

0

0

0

0

166.1

2.34

 

0

0

0

0

172.59

3.97

 

1

157.58

2.01

 

0

0

0

0

149.65

1.83

 

0

0

145.18

1.71

 

0

145.54

1.6

 

0.01

0

142.25

1.44

 

0

0

0

136.52

1.25

 

0.01

0

0

0

110.91

2.26

 

1.07

0

0

c

116.88

0.74

 

0.05

0

0

0.03

104.98

0.35

 

0.05

0

0.01

0.04

99.63

0.11

 

0.01

0.01

0

0.01

a.

The moisture level varies depending on the process details. The calculated values

b.

assume M = O.

The sulfur level varies depending on the process details. The calculated values

 

assume S = O.

c.Variable.

Values calculated from the North American Combustion Handbook, Tables 2.1a and 2.1b.

57

Image 57
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple OptionalConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixFilter Purging Calibration GAS CheckDust in the Flue GAS Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04