Teledyne 9060Z Function up and Function down keys, Option up and Option down keys, Enter key

Page 30

Model 9060Z Manual

FUNCTION up and FUNCTION down keys

These two keys allow for the selection of the setup function. A function summary table is found at the start of the relevant SETUP, COMMISSIONING or CALIBRATION mode chapter.

OPTION up and OPTION down keys

These two keys allow for changing the option for the selected function. A list/range of options for each function is found in the function summary table at the start of the relevant SETUP, COMMISSIONING or CALIBRATION mode chapter.

ENTER key

The ENTER key sets the currently displayed option and stores the value in non‐volatile memory to be retrieved on device start up. If the ENTER key is not pressed when a new option is chosen, the previous option will be retained.

Probe impedance key

When this key is pressed the transmitter will measure the impedance of oxygen sensor(s) attached to the transmitter. If the burner is not enabled (terminals 10 & 11) or the probe temperature is below 700°C (1292°F) impedance checking will not be performed. A ‘Z’ will be seen in the bottom left hand corner of the display as the test is performed.

Auto calibrate key

When this key is pressed the transmitter will calibrate the analogue output channels that are set to auto calibration (see chapter 10). This is performed by directing the output current away from the output terminals (terminals 12&13 and 14&15) back to an analogue input to the transmitter in order to calculate a zero and span calibration factor for each of the output channels.

During this process normal output to the analogue channels will be interrupted sending the outputs to zero mA.

30

Image 30
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Alarm relay contacts Temperature accuracyOutput load Alarm relays PowerConnection Weight Temperature RangeHead Temperature 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting 4B Earth Connection PE Installing the Auxiliary Thermocouple4A Shield Connections Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Alarms Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkProbe or Sensor Calibration Onnecting PowerCommissioning ‐ RUN Mode For more details see AppendixFilter Purging Calibration GAS CheckDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayOxygen Display Units Keypad in RUN modeKey text RUN mode White text Setup mode Black text KeypadKeypad in Setup / Commissioning / Calibration modes Enter key Function up and Function down keysOption up and Option down keys Probe impedance keyChanging Options Setup ModeFunction Summary Table Function name RangeSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Calibration Freezes Outputs Temperature UnitsFlue Pressure Units and Value Solenoid 1 & 2 OperationAlarm Relay 1, 2 and 3 Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Common Alarm Relay FunctionReference Air Pump Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Cleaning MaintenanceTransmitter Maintenance Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the N/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the S/C atom ratio in the fuelProbe EMF Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 26.2 13.120.7 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x03 0x41 Special Instruction Function0x02 0x04