Teledyne 9060Z manual Communications Port Options, Alarm Log Clearing

Page 41

Model 9060Z Manual

7.2.25 Communications Port Options

The 9060Z transmitter has a serial communications port available at terminals 18 to 22. The default protocol is for RS‐232 running at 19,200 baud rate with 8 bits, Even parity and 1 stop bit.

 

 

 

Function

 

 

Options

 

 

 

 

60

 

 

Serial Interface

 

 

RS‐232 or RS‐485 (Use RS‐485 for MODBUS)

 

 

 

61

 

 

Serial Baud Rate

 

9600, 19200, 38400, 57600 (RS‐232 up to 19200 only)

62

 

 

Serial Parity

 

 

Even, Odd or No Parity

63

 

 

MODBUS

 

Set the MODBUS address for this device between 1 and 246.

 

Address

 

0 is to disable the MODBUS

7.2.26 Alarm Log Clearing

Every alarm that is instigated internally in the transmitter or as a process level is recorded in the alarm log with the activation time, accepted time and the cleared time. The last 4000 events will be recorded and then the oldest alarms will be dropped off as new ones occur. If you would like to delete the log recording select CLEAR in SETUP step 64 by using the OPTION keys and press the ENTER key. The message “Alarm Log Cleared” will be displayed.

41

Image 41
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Output load Alarm relays Temperature accuracyAlarm relay contacts PowerHead Temperature Temperature RangeConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting 4A Shield Connections Installing the Auxiliary Thermocouple4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Output Channels Connecting the Auxiliary Thermocouple OptionalConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting the Transmitter to a Modbus Network Connecting Reference AIRCommissioning ‐ RUN Mode Onnecting PowerProbe or Sensor Calibration For more details see AppendixDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKey text RUN mode White text Setup mode Black text Keypad in RUN modeOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Option up and Option down keys Function up and Function down keysEnter key Probe impedance keyFunction Summary Table Setup ModeChanging Options Function name RangeLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Flue Pressure Units and Value Temperature UnitsCalibration Freezes Outputs Solenoid 1 & 2 OperationMaximum Calibration Gas 1 & 2 Positive / Negative Error Oxygen Content Calibration Gas 1Alarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionSelecting the Correct Fuel Operation of the Alarm Relays when an Alarm is AcceptedReference Air Pump Options Function OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Transmitter Maintenance MaintenanceCleaning Replacement PartsAuxiliary Thermocouple IndexTroubleshooting Is the O/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Appendix 2, Probe EMF TablesProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 20.7 13.126.2 33.8Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x02 0x41 Special Instruction Function0x03 0x04