19.‘Gas 1 Calibration Error’

20.‘Gas 2 Calibration Error’

At the end of an automatic gas calibration check cycle the oxygen reading was outside the values set in the COMMISSIONING menu #26 to 28 (#35 to 37). This alarm will not reset until the next calibration check cycle. The cycle can be initiated manually or automatically.

21. ‘BBRAM Fail’

All of the setup options are held in the battery backed memory (BBRAM). This is the battery shaped device at the bottom centre of the 1732‐1 PCB labeled MEM1. This alarm will occur when this device fails and will need to be replaced.

9.2 SELECTABLE PROCESS ALARMS

The alarm relays 1, 2 and 3 are generally used to transmit a gas related alarm event. Any or all of the following functions can be selected for each relay.

The trip levels and the delay times are set is in the COMMISSIONING menu.

NOTE: The process alarms will only be activated if they are enabled in the SETUP function 6. See chapter 6.3.5.

21.‘Oxygen 1 Very Low’

22.‘Oxygen 2 Very Low’

The measured oxygen level on the indicated probe has been below the trip level shown in COMMISSIONING menu function #38 for longer than the delay time shown in function #39.

23.‘Oxygen 1 Low’

24.‘Oxygen 2 Low’

The measured oxygen level on the indicated probe has been below the trip level shown in COMMISSIONING menu function #40 for longer than the delay time shown in function #41.

25.‘Oxygen 1 High’

26.‘Oxygen 2 High’

The measured oxygen level on the indicated probe has been above the trip level shown in COMMISSIONING menu function #42 for longer than the delay time shown in function #43.

27. ‘Oxygen Deviation’

The difference between the oxygen level measured on probe #1 and the oxygen level measured on probe #2 is greater than the trip level shown in COMMISSIONING menu function #44 for longer than the delay time shown in function #45.

28.‘Probe 1 Temperature Low’

29.‘Probe 2 Temperature Low’

The probe temperature is under 650°C (1200°F). The oxygen reading is therefore invalid. If the probe heater has been on for more than 20 minutes and the temperature is less than 650°C (1200°F) a ‘Probe # Heater Fail’ alarm will occur.

NOTE: The ‘Probe # Temperature Low’ relay function is used with unheated probes to indicate oxygen reading is invalid (the probe is below 650°C (1200°F) ), in case the process temperature falls below this

47

Image 47
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Series 9060H Oxygen Probes & Sensors Introduction9060Z Transmitter Model 9060Z Manual Power Temperature accuracyOutput load Alarm relays Alarm relay contacts9060H 9060UL/UH Application Temperature RangeHead Temperature Connection WeightHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumMounting the Transmitter Installation and CommissioningInstalling a 9060H Oxygen Probe Oxygen Probe Mounting Electrical Connections Installing the Auxiliary Thermocouple4A Shield Connections 4B Earth Connection PEModel 9060Z Manual Model 9060Z Manual Heater Supply Interlock Connection for Heated Probes Connecting AN Oxygen Probe CableHeater Interlock Relays Model 9060Z Manual Connecting the Automatic Purge and Calibration Check System Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the AlarmsConnecting the Transmitter to a Modbus Network Connecting Reference AIRFor more details see Appendix Onnecting PowerCommissioning ‐ RUN Mode Probe or Sensor CalibrationDust in the Flue GAS Filter PurgingCalibration GAS Check Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKeypad Keypad in RUN modeKey text RUN mode White text Setup mode Black text Oxygen Display UnitsKeypad in Setup / Commissioning / Calibration modes Probe impedance key Function up and Function down keysOption up and Option down keys Enter keyFunction name Range Setup ModeFunction Summary Table Changing OptionsLower Line Items Setup Mode FunctionsProbe 1 Offset Damping Factor Process AlarmsOxygen Display Units Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Solenoid 1 & 2 Operation Temperature UnitsFlue Pressure Units and Value Calibration Freezes OutputsCommon Alarm Relay Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Alarm Relay 1, 2 and 3 FunctionFunction Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Reference Air Pump OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationFunction # Function Description Options GAS Calibration Check and PurgeCalibration GAS Software Upgrades First Approach TroubleshootingDetailed Fault Analysis Replacement Parts MaintenanceTransmitter Maintenance CleaningAuxiliary Thermocouple IndexTroubleshooting Is the S/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the N/C atom ratio in the fuelProbe EMF @ 720C 1320F Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMFZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 33.8 13.120.7 26.2Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x04 0x41 Special Instruction Function0x02 0x03