Teledyne 9060Z manual Onnecting Power, Commissioning ‐ RUN Mode, Probe or Sensor Calibration

Page 24

Model 9060Z Manual

network line from the transmitters is taken from the middle of the transmitter network string, a terminating resistor should be enabled with LK3 at each end of the network line.

The MODBUS™ protocol of the network is –

Baud Rate

19,200

Parity

Even

Stop Bits

1

RS485

2‐wire (half duplex)

Mode

RTU (binary mode)

For more details see Appendix 5.

4.14 ONNECTING POWER

Before commissioning the probe, sensor or transmitter, read the CAUTION paragraphs at the front of this manual. The green & yellow EARTH wire MUST be connected to earth.

4.15 COMMISSIONING ‐ RUN MODE

When the transmitter is turned on it will go to RUN mode. The SET‐UP/RUN button will toggle between the two modes. The upper line of the display will now read ‘% OXYGEN’. If the probe or sensor temperature is not above 650°C (1200°F), a “Probe Low Temperature” message is flashed on the lower line. The probe or sensor temperature can be checked on the lower line of the display.

4.16 PROBE OR SENSOR CALIBRATION

The zirconia sensor provides an absolute measurement of oxygen partial pressure. There are no calibration adjustments, apart from ‘Probe 1 Offset’, for the probe. The zirconia sensor EMF is either correct or it needs to be replaced. To check that the probe is functioning correctly, first check that the

24

Image 24
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Forcing a Cold Start Resetting the Calibration Factors PurgeEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Temperature accuracy Output load Alarm relaysAlarm relay contacts PowerTemperature Range Head TemperatureConnection Weight 9060H 9060UL/UH ApplicationHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Function Minimum Maximum Range of outputsInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting Installing the Auxiliary Thermocouple 4A Shield Connections4B Earth Connection PE Electrical ConnectionsModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Auxiliary Thermocouple Optional Connecting the Output ChannelsConnecting the Alarms Connecting the Automatic Purge and Calibration Check SystemConnecting Reference AIR Connecting the Transmitter to a Modbus NetworkOnnecting Power Commissioning ‐ RUN ModeProbe or Sensor Calibration For more details see AppendixFilter Purging Calibration GAS CheckDust in the Flue GAS Connecting a Pressure Transducer StratificationDisplay and Keypad Graphical DisplayKeypad in RUN mode Key text RUN mode White text Setup mode Black textOxygen Display Units KeypadKeypad in Setup / Commissioning / Calibration modes Function up and Function down keys Option up and Option down keysEnter key Probe impedance keySetup Mode Function Summary TableChanging Options Function name RangeSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Temperature Units Flue Pressure Units and ValueCalibration Freezes Outputs Solenoid 1 & 2 OperationOxygen Content Calibration Gas 1 Maximum Calibration Gas 1 & 2 Positive / Negative ErrorAlarm Relay 1, 2 and 3 Function Common Alarm Relay FunctionOperation of the Alarm Relays when an Alarm is Accepted Selecting the Correct FuelReference Air Pump Options Function OptionsAlarm Log Clearing Communications Port OptionsCalibration Mode Menu Function name RangeCalibration Mode Functions Reference VoltagesHeater SSR Fault Correction Alarms Common AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Instrument Calibration Calibration SummaryProbe Calibration Cold StartGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Maintenance Transmitter MaintenanceCleaning Replacement PartsIndex Auxiliary ThermocoupleTroubleshooting Is the H/C atom ratio in the fuel Is the O/C atom ratio in the fuelIs the N/C atom ratio in the fuel Is the S/C atom ratio in the fuelAppendix 2, Probe EMF Tables Zirconia Oxygen Sensor Output mV Probe Type 9060H, SensorProbe EMF Probe EMF @ 720C 1320FZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 13.1 20.726.2 33.8Serial Configuration Modbus ProtocolModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x41 Special Instruction Function 0x020x03 0x04