7.2.16 Solenoid 1 & 2 Post Freeze

Duration of time the transmitter waits before resuming live readings following a gas calibration or purge event. (See also chapter 7.2.10)

7.2.17 Oxygen Content Calibration Gas 1 & 2

When configured as a gas calibration check solenoid this value specifies the oxygen content of the calibration gas.

7.2.18 Maximum Calibration Gas 1 & 2 Positive / Negative Error

Specifies the maximum positive & negative error levels used during the gas calibration check.

If the difference between the specified oxygen content and the measured process gas exceed these thresholds, a gas calibration error alarm will be triggered immediately.

7.2.19 Process Alarms

The oxygen measurements are continuously monitored by the transmitter for process alarm conditions (Process alarms must be enabled in the Set up menu). The limits and the alarm time delays are configurable for the Very Low Oxygen alarm, Low Oxygen alarm, High Oxygen alarm and the Oxygen Deviation alarm.

7.2.20 Alarm Relay 1, 2 and 3 Function

There are 3 user configurable alarm relays. Any of the 3 relays can be configured to be triggered on the following alarm conditions. In addition, any of the alarm conditions that are disabled from the common alarm relay can also be configured to trigger these 3 relays. Multiple selections can be made.

 

Process Alarms

Oxygen 1 Low

Probe 1 Temperature Low

Oxygen 2 Low **

Probe 2 Temperature Low **

Oxygen 1 Very Low

Calibration 1 In Progress

Oxygen 2 Very Low ** Calibration 2 In Progress **

Oxygen 1 High

Purge 1 In Progress

Oxygen 2 High **

Purge 2 In Progress **

Oxygen Deviation

 

7.2.21 Common Alarm Relay Function

The common alarm relay can be configured to be triggered on any of the following:

Instrument Alarms

Probe 1 Heater Fail

Reference Air Pump Fail

Probe 2 Heater Fail **

Reference Air Pump Overload

Probe 1 High Impedance

Alarm Log Fail

Probe 2 High Impedance **

Output 1 Fail

Probe 1 Thermocouple Open Circuit

Output 2 Fail

Probe 2 Thermocouple Open Circuit **

Heater 1 SSR Fail

Auxiliary Thermocouple Open Circuit

Heater 2 SSR Fail

** Are only available in dual probe mode.

Heater SSR Leakage

Probe 1 Filter Blocked Probe 2 Filter Blocked **

Gas 1 Calibration Error Gas 2 Calibration Error **

39

Image 39
Contents Teledyne Analytical Instruments Model 9060Z Manual Keypad Purge Forcing a Cold Start Resetting the Calibration FactorsEmail asktai@teledyne.com Important Notice Regarding 9060 Probe Option ‐ FIL‐3 Important Notices Introduction 9060Z TransmitterSeries 9060H Oxygen Probes & Sensors Model 9060Z Manual Power Temperature accuracyOutput load Alarm relays Alarm relay contacts9060H 9060UL/UH Application Temperature RangeHead Temperature Connection WeightHeated probes‐temperature range 0‐900C 1650F Model 9060Z Manual Range of outputs Function Minimum MaximumInstallation and Commissioning Installing a 9060H Oxygen ProbeMounting the Transmitter Oxygen Probe Mounting Electrical Connections Installing the Auxiliary Thermocouple4A Shield Connections 4B Earth Connection PEModel 9060Z Manual Model 9060Z Manual Connecting AN Oxygen Probe Cable Heater Interlock RelaysHeater Supply Interlock Connection for Heated Probes Model 9060Z Manual Connecting the Automatic Purge and Calibration Check System Connecting the Auxiliary Thermocouple OptionalConnecting the Output Channels Connecting the AlarmsConnecting the Transmitter to a Modbus Network Connecting Reference AIRFor more details see Appendix Onnecting PowerCommissioning ‐ RUN Mode Probe or Sensor CalibrationFilter Purging Calibration GAS CheckDust in the Flue GAS Stratification Connecting a Pressure TransducerGraphical Display Display and KeypadKeypad Keypad in RUN modeKey text RUN mode White text Setup mode Black text Oxygen Display UnitsKeypad in Setup / Commissioning / Calibration modes Probe impedance key Function up and Function down keysOption up and Option down keys Enter keyFunction name Range Setup ModeFunction Summary Table Changing OptionsSetup Mode Functions Probe 1 OffsetLower Line Items Process Alarms Oxygen Display UnitsDamping Factor Commissioning Mode Model 9060Z Manual Commissioning Mode Functions Model 9060Z Manual Solenoid 1 & 2 Operation Temperature UnitsFlue Pressure Units and Value Calibration Freezes OutputsCommon Alarm Relay Function Oxygen Content Calibration Gas 1Maximum Calibration Gas 1 & 2 Positive / Negative Error Alarm Relay 1, 2 and 3 FunctionFunction Options Operation of the Alarm Relays when an Alarm is AcceptedSelecting the Correct Fuel Reference Air Pump OptionsCommunications Port Options Alarm Log ClearingMenu Function name Range Calibration ModeReference Voltages Calibration Mode FunctionsHeater SSR Fault Correction Common Alarms AlarmsModel 9060Z Manual Selectable Process Alarms Alarm Relay Options Calibration Summary Instrument CalibrationCold Start Probe CalibrationGAS Calibration Check and Purge Calibration GASFunction # Function Description Options Software Upgrades Troubleshooting Detailed Fault AnalysisFirst Approach Replacement Parts MaintenanceTransmitter Maintenance CleaningAuxiliary Thermocouple IndexTroubleshooting Is the S/C atom ratio in the fuel Is the H/C atom ratio in the fuelIs the O/C atom ratio in the fuel Is the N/C atom ratio in the fuelProbe EMF @ 720C 1320F Appendix 2, Probe EMF TablesZirconia Oxygen Sensor Output mV Probe Type 9060H, Sensor Probe EMFZirconia Oxygen Probe Output mV Probe Type 9060UL/UH 33.8 13.120.7 26.2Modbus Protocol Serial ConfigurationModel 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual Model 9060Z Manual 0x04 0x41 Special Instruction Function0x02 0x03