Skew DAC Adjustment Procedure
Description
This procedure compensates for time delays between the AFG’s two DACs.
The skew setting which produces the lowest se co nd ha r monic amplitude is
found and loaded into non-volatile memory.
Equipment Setup
Connect the equipment as shown in Figure 3-3
Set up the Spectrum Analyzer:
Center Freque ncy = 8 MHz
Frequency Span = 3.2 kHz
Adjustment Procedure
1. Reset the AFG:
*RST
Figure 3-3. Skew DAC Adjustment Setup
Agilent E1445A Service Manual Adjustments 109