Functional Verification: Output Relay Test
Description
The purpose of this test is to check the output relay.
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2. Set up equipment as shown in Figure 2-4:
3. Set up the A FG to output a 1 MHz sinewave:
FREQ 1E6; Set freq to 1 MHz
:VOLT 4VPP Set AFG amplitude
INIT:IMM Initiate
4. Verify that a 1 MHz sinewave appears on the scope.
5. Disable the Output relay:
OUTP OFF
6. Verify that no signal appea rs on the scope.
Figure 2-4. Output Relay Test Setup

Agilent E1445A Service Manual Verification Tests 25