Table 2-11. Performance Test Record for the Agilent E1445A (Pag e 3 of 7)

Model _____________________________ Report No. ________ ____ ________________ Date _______________
Test
Description* Minimum Measured
Reading Maximum Meas
Uncert TAR
Test 2-1. DC Zeros Test (Values in Vdc)
No Filter:
10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB at ten)
2.0426V (14 dB at ten)
0.3238V (30 dB at ten)
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044
____________
____________
____________
____________
____________
____________
____________
____________
____________
0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
250 kHz Filter:
10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB at ten)
2.0426V (14 dB at ten)
0.3238V (30 dB at ten)
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044
____________
____________
____________
____________
____________
____________
____________
____________
____________
0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
10 MHz Filter:
10.2375V (0 dB atten)
9.1347V (.99 dB atten)
9.1241V (1 dB atten)
8.1319V (2 dB atten)
6.4594V (4 dB atten)
4.0756V (8 dB atten)
2.2918V (13 dB at ten)
2.0426V (14 dB at ten)
0.3238V (30 dB at ten)
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.022
-0.0044
-0.0044
____________
____________
____________
____________
____________
____________
____________
____________
____________
0.022
0.022
0.022
0.022
0.022
0.022
0.022
0.0044
0.0044
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
1E-6
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1

* Since the arbitrary waveform consists of zeros, the expected reading is 0 V, regardless of the amplitude

setting. The amplitude is changed in order to turn the vari ous a tte nuators on and off.

Agilent E1445A Service Manual Verification Tests 77