Test 2-2: DC Accuracy (cont’d)
Test Procedure
(cont’d)
Table 2-2. DC Accuracy Test Points
Amplitude
(volts) Filter Test Limits
(volts)
10.2375
5.0
0.0
-5.0
-10.24
10.2375
-10.24
10.2375
-10.24
None
None
None
None
None
250 kHz
250 kHz
10 MHz
10 MHz
10.2375 ± 0.0512
5.0 ± 0.0355
0.0 ± 0.0205
-5.0 ± 0.0355
-10.24 ± 0.0512
10.2375 ± 0.0512
-10.24 ± 0.0512
10.2375 ± 0.0512
10.24 ± 0.0512
Example Program

This program performs the DC Accuracy test.

10! RE-STORE "DC_LEVELS"
20 DIM Vout(1:9),Filter(1:9 )
30 !
40 !---------- Set up I/O path and reset AFG -----------
50 ASSIGN @Afg TO 70910
60 OUTPUT @Afg;"*RST;*CLS" !Reset AFG
70 !
80 !---------- Ini tialize variables ----------
90 DATA 10.2375,5.0,0,-5.0,-10.24,10.2375,-10.24 ,10. 2375,-10.24
100 READ Vout(*)
110 !
120 DATA 0, 0,0,0,0,1,1,2,2
130 READ Filter(*)
140 !
150 !---------- Set u p D MM ----------
160 CLEAR SCREEN
170 PRINT "Set up DMM:"
180 PRINT
190 PRINT " Function -- DCV"
200 PRINT " Range -- AUTO"
210 PRINT
220 PRINT "Connect DMM HI and LO to AFG Outp ut."
230 DISP "Press ’Continue’ when ready"
240 PAUSE
250 CLEAR SCREEN

Agilent E1445A Service Manual Verification Tests 39