Functional Verification: Start Arm In Test
Description
The purpose of this test is to ch eck the Start Arm In port. The "TRIG OUT"
port of the Command Module is used to sen d a Sta r t A rm signal to the AFG.
Test Procedure
1. Reset the AFG:
*RST;*CLS Reset AFG and clear
status registers
2. Set up equipment as shown in Figure 2-2:
3. Send the following commands to the Command Module to output 0 V
to the "Trig Out" port:
*RST
OUTP:EXT:STAT ON
OUTP:EXT:SOUR INT
OUTP:EXT:LEV 1
Figure 2-2. Start Arm In Test Setup