Calibration Commands (cont’d)
CALibration:DATA:FILTer <block> transfers the two calibration
constants that are used to determine the fr eq u e nc y points that will be
calibrated for the 10 MHz filter. The q uery form returns the current
constants in IEEE-488.2 definite block data format. See the AC
Flatness Adjustment procedure for the 10 MHz filter for more
information on the use of this command.
CALibration:DATA:SKEW <data> transfers the calibration
constant that is used by the skew DAC to synchronize the AFG’s
DAC’s. The query form returns the current constant in IEEE-488.2
definite block data format. See the Skew DAC Adjustment procedur e
for more information on the use of this command.
NOTE The CAL:DATA:FILT and CAL:DATA:SKEW commands are available only
on units with firmware rev A.02.00 or higher (use the *IDN? command to
determine the AFG’s firmware revision).
CALibration:STATe:AC <state> enables or disables A C
corrections using the calibration constants. If <state> is ON,
corrections will be used. If <state> is OFF, corrections will not be
used.
CALibration:STATe:DC <state> enables or disables D C
corrections using the calibration constants. If <state> is ON,
corrections will be used. If <state> is OFF, corrections will not be
used.
CALibration:STATe <state> enables or disables both AC and DC
corrections using the calibration constants. If <state> is ON,
corrections will be used. If <state> is OFF, corrections will not be
used.
*PUD <data>stores the specified data in non-volati le "pr ote c te d
user data" memory. The data must be sent in IEEE-488.2 definite or
indefinite block format. The query form (*PUD?) returns the
current protected user data in IEEE-488.2 definite block format.

Agilent E1445A Service Manual Adjustments 85