Start When Stable – Taring continues until the

Sample Tare – Records the average initial sample weight during a programmable period of time. After lid closure, there is a 7 second delay before the sample tare time begins. There are five (5) possible choices under this menu. The factory default is Start After 8s.

Start When Stable – Taring continues until the

weight is found stable (≤ 0.2mg deviation) over 4s of time.

Start After (selectable) Seconds: Test starts after (selectable) seconds of averaging, i.e. 3s, 5s, 8s, or 20s

Special Tare - There are three special tare options. Two of these set up operations that are not required for most types of testing.

Normal Tare – selects normal operation of tests.

Syringe Tare – Used to obtain an accurate sample tare for materials

that rapidly evaporate during the sample loading process. An external balance with 0.0001-gram (0.1 mg) resolution is used to weigh the syringe before and after loading, which are entered into the MAX 5000XL. The difference between the loaded and empty syringe weights is start weight.

Foil Tare – is used when a sample of sprayed material, such as paint, is to be tested. Up to 50 pieces of foil can be pre-weighed before actual testing begins. The alphanumeric identity of each foil is stored in memory until it is replaced by other foil tare entries. During a test, the ID of the foil is entered from a menu of stored foils. The weight of the foil is subtracted from the measured weight, giving the true weight of the sprayed sample. (See section 6.5 for more information.)