Sample Tare – Records the average initial sample weight during a programmable period of time. After lid closure, there is a 7 second delay before the sample tare time begins. There are five (5) possible choices under this menu. The factory default is Start After 8s.
Start When Stable – Taring continues until the
weight is found stable (≤ 0.2mg deviation) over 4s of time.
Start After (selectable) Seconds: Test starts after (selectable) seconds of averaging, i.e. 3s, 5s, 8s, or 20s
Special Tare - There are three special tare options. Two of these set up operations that are not required for most types of testing.
Normal Tare – selects normal operation of tests.
Syringe Tare – Used to obtain an accurate sample tare for materials
that rapidly evaporate during the sample loading process. An external balance with
Foil Tare – is used when a sample of sprayed material, such as paint, is to be tested. Up to 50 pieces of foil can be