No heat (cont.) | Door sensor failure | Check display for LID OPEN | |
|
| message when open during a | |
|
| test, call AZI at | |
|
| or | |
|
| assistance | |
|
|
| |
| Temperature setpoint too low | Raise temperature above | |
|
| room ambient or standby | |
|
| temperature, see pages 25 & | |
|
| 68 | |
|
|
| |
No power to | Power switch off | Check switch position, | |
instrument |
| confirm power cord seating, | |
Power cord disconnected | |||
| remove and test fuse, see | ||
|
| ||
| Fuse blown | ||
| page 14, test the outlet with a | ||
|
| ||
| Outlet not powered | lamp or meter. | |
|
|
| |
No weight reading | Balance not initialized | Turn power off, wait 5 | |
|
| seconds, then turn power on, | |
|
| wait 15 seconds. Call AZI at | |
|
| ||
|
| ||
|
|
| |
Erratic results | Poor AC ground | Ensure instrument is | |
|
| grounded from case ground | |
|
| to earth ground, Call AZI at | |
|
| ||
|
| ||
|
|
| |
| Dirt in chamber | Clean unit, | |
|
| see pages 78 to 82 | |
|
|
| |
| Dirt in pan support | See page 78 or call AZI at | |
|
| ||
|
| ||
|
|
| |
Display erratic | Unit cold (near freezing) | Allow to | |
|
|
| |
THE BALANCE | Unit powered ON without pan | Power off instrument; Install | |
REPORTS | support in place. | pan support and power on | |
UNDERLOAD… |
| instrument, see page 13 | |
|
|
| |
UNDER appears on | Pan support has been | Power off instrument; Install | |
display when | removed since instrument was | pan support and power on | |
[START] is pressed | powered ON | instrument, see page 13 | |
|
|
| |
Instrument won't tare | Test chamber temperature | Allow system to stabilize. | |
| fluctuates/changes |
| |
| Check tare options, | ||
|
| ||
|
| see page 30 | |
|
|
| |
| Dirt in test chamber or under | Clean instrument, see page | |
| pan support | 78 to 82 and following | |
|
|
|