 10. ROUTINE MAINTENANCE

 10.1. Chamber InteriorManual backgroundManual background 10. ROUTINE MAINTENANCE

The MAX 5000XL requires routine cleaning to keep the test chamber free of excess Manual background sample material. Failure to keep the instrument interior clean reduces its accuracy Manual background and repeatability. Sample material collecting in the test chamber prevents free Manual background movement of the pan support. In addition, sample material collecting on the rim of Manual background the heater shroud prevents the lid from closing completely, causing erratic results. Manual background Material may fall into the system's interior, causing balance problems. Routine Manual background cleaning of the test chamber reduces the likelihood of problems and repairs, Manual background extending the life of the instrument.Manual background

Manual backgroundManual background 10.1. Chamber Interior

The chamber interior should be cleaned when needed to eliminate the buildup of sample material. To clean debris from the chamber’s

interior:

Switch the main power switch OFF.

Use a vacuum line, portable vacuum cleaner, or a small paintbrush to remove debris from the area around the balance cover and pan support stem.

A compressed air line or canned air is not recommended since debris migrates into the system's interior and settles in the electronic components and balance stem.

Do not clean the shroud or heater element with a brush, vacuum, cleaner liquid, or any object. Use the “Oven Chamber Cleaning” procedure.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 78 of 92