 4.7. Calibrate the Balance - 50 gram weight

Manual backgroundManual background 4.7. Calibrate the Balance - 50 gram weight

NOTE: The MAX 5000XL must be powered on until the balance temperature stability icon in the lower left corner of the main screen changes from unstable Manual background to stable Manual background to allow the balance to stabilize before beginning the balance calibration. (See section 6.1 Balance Temperature Stability on page 64 for more information). The unit should not be moved, jarred, or otherwise disturbed in any way during the calibration period. A vibration-free table will aid the calibration process. If vibration is excessive, the instrument will stop the calibration process and indicate the reason.

The balance should be calibrated:

Immediately after initial setup and warm-up,

Whenever the instrument is moved,

Whenever a problem is suspected, or

As your quality policy dictates.

From the Main Test Screen, press

[Menu].

Select the CALIBRATION MENU.

Select BALANCE CALIBRATION.

Select CALIBRATE BALANCE.

The first screen reminds the operator that the instrument must be free of vibration during the calibration and that a 50-gram calibration weight must be available.

Press [GO].

If the chamber temperature is too hot, the screen will prompt the operator to

“OPEN LID CHAMBER TEMPERATURE IS TOO HOT WAIT FOR COOL DOWN”

If the chamber temperature is cool enough to begin the calibration, one of two prompts will appear:

If the lid is closed, the “OPEN LID”

“PLACE CLEAN PAN ON PAN

SUPPORT” prompt will appear.

If the lid is open, the “PLACE CLEAN

PAN ON PAN SUPPORT CLOSE LID”

prompt will appear.