4.7. Calibrate the Balance - 50 gram weight
NOTE: The MAX 5000XL must be powered on until the balance temperature stability icon in the lower left corner of the main screen changes from unstable to stable to allow the balance to stabilize before beginning the balance calibration. (See section 6.1 Balance Temperature Stability on page 64 for more information). The unit should not be moved, jarred, or otherwise disturbed in any way during the calibration period. A
The balance should be calibrated:
∙Immediately after initial setup and
∙Whenever the instrument is moved,
∙Whenever a problem is suspected, or
∙As your quality policy dictates.
∙From the Main Test Screen, press
[Menu].
Select the CALIBRATION MENU.
Select BALANCE CALIBRATION.
Select CALIBRATE BALANCE.
The first screen reminds the operator that the instrument must be free of vibration during the calibration and that a
∙Press [GO].
If the chamber temperature is too hot, the screen will prompt the operator to
“OPEN LID CHAMBER TEMPERATURE IS TOO HOT WAIT FOR COOL DOWN”
If the chamber temperature is cool enough to begin the calibration, one of two prompts will appear:
∙If the lid is closed, the “OPEN LID”
“PLACE CLEAN PAN ON PAN
SUPPORT” prompt will appear.
∙If the lid is open, the “PLACE CLEAN
PAN ON PAN SUPPORT CLOSE LID”
prompt will appear.
AZI Customer Service | Page 20 of 92 |