 Special Features

 5.5. Sample Foils Enter Foils Special Features

This menu provides optional features for the instrument. A Web Server and Parameter development tools are examples of additional features that can be purchased. If you have purchased one of these features, you have been given an access code to enable/disable the feature. Entering the code will toggle the feature on/off. For more information regarding the special features available on your MAX 5000XL, please call your technical sales representative or AZI Customer Service at 800-528-7411 or 602-470-1414.

Manual backgroundManual background 5.5. Sample Foils

Selecting this option from the Main Menu will begin the process of weighing and storing the weights of 50 foils. The foils are used in paint processes where the paint is sprayed onto the foil as the end product is being painted. When placed into the MAX 5000XL, the weight of the foil will be subtracted from both the start weight and ending weight so that only the spray product will be considered in the moisture determination. Pan tare will first tare the empty sample pan, see pg. 30.

Manual backgroundManual background Enter Foils

From Main Menu, select Sample Foils. Select [ADD] or [DELETE] to add or delete a foil.

AZI Customer Service 800-528-7411 or 602-470-1414