Temperature Calibration
The MAX 5000XL temperature calibration routines will calibrate and verify the accuracy of the oven temperature throughout the usable range. The oven is calibrated at twelve
(12)preset increments and verified at temperatures that are user selectable.
Verification Setup
The VERIFICATION SETUP menu allows the verification of the temperature calibration to be enabled or disabled, and selection of which temperatures are to be verified. Selecting ALL CAL POINTS will cause verification of twelve (12) preset temperatures from 50º C to 575º C. Four (4) custom temperatures may be entered to verify only those temperatures used in your processes. Entering a number above 575º C will produce a warning message on the display and set in the default highest
temperature of 575º C. Press [OK] to remove the warning from the display.
Calibration Module Setup
This menu allows entry of the Temperature Calibration Interface (TCI) data. Serial numbers of the interface module and probe used for the calibration and the certification date of the TCI is entered here and saved for printing in the various reports.