After placing a clean pan on the balance, close the lid. The screen will
display “CALIBRATING ZERO LOAD,
PLEASE WAIT.”
As soon as the tare is finished, the next prompt appears.
∙Open the lid and gently place a
Avoid excess pressure on the balance.
Use the tweezers included with the MAX 5000XL to gently place the weight on the pan.
As soon as the instrument detects weight on the balance, the prompt “CLOSE LID” appears.
With the lid closed, the prompt “CALIBRATING, PLEASE WAIT” appears.
At the conclusion of the calibration one of two prompts will appear, indicating the PASS/FAIL condition of the calibration.
∙If the result is “PASS”, remove the weights and the instrument is ready for operation.
∙Press [PROCEED] to return to the BALANCE CALIBRATION menu.
∙Press [ESC] three (3) times to return to the Main Test Screen
∙If the result is “FAIL”, the calibration was unsuccessful. Rerun the calibration.
∙If the calibration fails again, contact AZI Customer Service at