Arizona MAX-5000XL user manual The Precision Weight Test

Models: MAX-5000XL

1 92
Download 92 pages 19.89 Kb
Page 71
Image 71
 8.2. The Precision Weight Test

Manual backgroundManual background 8.2. The Precision Weight Test

The precision weight test consists of performing the weight test in Section 8.1 multiple times to determine the instruments repeatability. This is a more critical verification of the instrument’s performance.

Mean, Standard Deviation, and Relative Standard Deviation are calculated using the following formulas.

Mean = total of results divided by number of results.

Standard Deviation =

Σ(

x

- x)2

n - 1

 

Relative Standard Deviation expressed as percentage = Standard Deviation divided by Mean, multiplied by 100

Perform the steps in the weight test a total of five (5) times.

After all five tests are complete, press [MENU] and select the

STORED DATA MENU.

PRESS [ENT] to select tests.

Arrow up or down to select the five weight tests and press [ANALYZE].

The mean of the results should be 37.500%, +/-.020 for a test using the five and three gram weights.

If the results obtained from these tests are outside the specified range, call AZI Customer Service at 800-528-7411 or 602-470-1414 for assistance.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 71 of 92

Page 71
Image 71
Arizona MAX-5000XL user manual The Precision Weight Test