Arizona MAX-5000XL Hardware, The MAX 5000XL is designed to remain powered up at all times, Oven

Models: MAX-5000XL

1 92
Download 92 pages 19.89 Kb
Page 10
Image 10
 3.1. Hardware

The MAX 5000XL is designed to remain powered up at all times. 3.2. Oven 3.1. Hardware

The MAX 5000XL consists of:

A ceramic convection air oven

A precision balance weighing mechanism with a resolution of 0.0001 gram and a full scale range of 100 grams

A backlit, color graphic liquid crystal display

A microprocessor that controls the test process, data handling, communications, and performs all necessary calculations, and

Connectors on the back panel for an Ethernet connection, parallel printer, and serial communications devices (Other connections are provided that will be implemented in later revisions to the instrument.)

The MAX 5000XL is designed to remain powered up at all times.

The balance requires a lengthy warm up period. The first test of the day can be made sooner and with more accuracy if the unit remains powered over night. When not in use, select a test with start temperature of 25º C and keep the lid down. The oven will remain cool and the Liquid Crystal Display (LCD) backlight will turn off after a period of inactivity to conserve energy. Press any key to restore illumination to the LCD.

If the test process is interrupted by outside interference such as static discharge, surges or drops in line voltage, or otherwise freezes its operation and/or display, reset the instrument by recycling power. By turning the instrument off and back on, the instrument should return to normal operation. If the instrument does not return to normal operation, contact AZI Customer Service at 800-528-7411, 602-470-1414, or by e-mail at support@azic.com.

Manual backgroundManual background 3.2. Oven

The ceramic sample oven maintains accurate temperature control to prevent sample degradation as it removes volatile components as quickly as possible. A Resistance Temperature Device (RTD) mounted near the heater measures oven temperature. Temperature is held within one or two degrees of the set value (depending on the set value) after equilibrium is reached. The oven usually reaches test temperature in 90 to 180 seconds, depending on the difference between the starting and setpoint temperatures. The oven is automatically turned off when the test is completed.

Due to the nature of the oven lining, the oven is cool almost immediately after opening the lid. The sample will remain hot much longer than the oven itself so caution is required when removing used sample pans from the oven.

AZI Customer Service 800-528-7411 or 602-470-1414