 1. WARNINGS AND SAFETY INFORMATION

Lithium Coin Cell BatteryPrevent excess heat build-up around the analyzerManual background 1. WARNINGS AND SAFETY INFORMATION

The MAX 5000XL moisture analyzer complies with the Underwriter Laboratories Inc. standards and European Council Directives, for Electrical Equipment for Laboratory Use, electromagnetic

compatibility, and the stipulated safety requirements. However, improper use or handling of the instrument can result in damage to the equipment and/or injury to personnel.

The MAX 5000XL is Equipment Class I, Pollution Degree 2, and Installation Category II.

Read these operating instructions thoroughly before using your moisture analyzer to prevent damage to the equipment. Keep these instructions in a safe place.

The following instructions will help to ensure safe and trouble-free operation of your MAX 5000XL.

Use the analyzer only for performing moisture/solids/ashing analysis of samples. Any improper use of the analyzer can endanger persons and may result in damage to the analyzer or other material assets.

Place analyzer in an appropriate fume hood for ashing and/or burning a sample.

When ashing and/or burning a sample, use the ash rate to slow ignition and prevent flaming. If the sample flames, it will damage the heater and produce poor test results.

Do not use this analyzer in a hazardous area/location; operate it only under the ambient conditions specified in these instructions.

Although the analyzer is simple to operate and is very user friendly, it should only be operated by qualified persons who are familiar with the properties of the sample being analyzed.

Ensure, before getting started, that the voltage rating printed on the manufacturer's label is identical to your local line voltage (see the section on AC Power Connection beginning on page 14).

The MAX 5000XL is provided with a power cord that has a protective grounding conductor. Do not operate the unit without the supplied three-wire power cord or an identical recognized equivalent that meets all applicable standards.

AZI does not recommend the use of an extension cord. However, if one must be used, use an extension cord that meets all applicable standards and has a protective grounding conductor.

Do not remove the protective grounding wire from the power supply circuit

With the power switch off, power is still applied to the power entry module. Remove the power cord from the MAX 5000XL to ensure complete power removal from the instrument.

Position the power cable so that it cannot touch any hot areas of the analyzer.

Use only AZI supplied accessories and options with the MAX 5000XL or check with AZI before using any third party accessories, such as printers and analytical balances, to verify compatibility.

If there is visible damage to the MAX 5000XL or its power cord, unplug and isolate the unit first and then call Customer Service at 800-528-7411 or 602-470-1414.

Do not open the housing of the MAX 5000XL. There are no user serviceable parts or adjustments inside the unit. Any maintenance inside the unit is to be performed by

factory-trained technicians only. Any unauthorized inspection and/or maintenance of the MAX 5000XL will make the warranty null and void.

Lithium Coin Cell Battery

The MAX 5000XL contains a lithium battery to keep power on the clock when the unit is turned off.

CAUTION – The battery may explode if mistreated. Do not recharge, disassemble or dispose of in fire.

Prevent excess heat build-up around the analyzer

When setting up the analyzer, leave enough space to prevent heat from building up and to keep your analyzer from overheating. Leave 20 cm (about 8 in.) around the analyzer and 1 m (about 3 ft.) above the unit.

Do not put any flammable substances on, under or near the analyzer, because the area around the unit will heat up, especially during the self-cleaning cycle.

Use extreme care when removing the sample from the chamber. Although the ceramic heating chamber cools rapidly, the sample itself, the sample pan, and the heating element can still be extremely hot.

AZI Customer Service 800-528-7411 or 602-470-1414