View Calibration Report

As soon as the instrument detects weight on the balance, the prompt “CLOSE LID” appears.

With the lid closed the prompt “CALIBRATING, PLEASE WAIT” appears.

At the conclusion of the calibration, the “OPEN LID” “REMOVE WEIGHT” prompt appears with a PASS/FAIL indication of the calibration success.

If the result is “PASS”, open the lid; remove the weight and press [PROCEED] to continue.

If the result is “FAIL”, the calibration was unsuccessful. Rerun the calibration.

If the calibration fails again, contact AZI Customer Service at 800-528-7411 or 602-470-1414.

Press the ESC key three (3) times to return to the Main Test Manu.

View Calibration Report

When VIEW CALIBRATION REPORT is selected, the last balance calibration date will be displayed.

Calibration Reminder Setup

A prompt is available to remind you when scheduled calibrations are due. The default factory settings are:

ENABLE - No

TIME – 00:00 INTERVAL - 0 Weeks

To turn the prompt on, off, or adjust the interval; go to the Main Test Screen and select in order: MENU, CALIBRATION MENU, BALANCE CALIBRATION, and CALIBRATION SETUP REMINDER

Enable – Toggle to YES to enable the prompt or NO to prevent the prompt from appearing.

Time – The time the prompt is to appear, usually the start of the shift (using a 24 hour format).

Interval – The interval between calibrations (in weeks). If the interval is set to

“00”, the reminder will occur daily.

AZI Customer Service 800-528-7411 or 602-470-1414