Arizona MAX-5000XL user manual Shroud Gasket, Insulator Seal

Models: MAX-5000XL

1 92
Download 92 pages 19.89 Kb
Page 79
Image 79
 10.2. Shroud Gasket

 10.3. Insulator SealManual background 10.2. Shroud Gasket

The shroud gasket creates a seal between the heater shroud and ceramic plate. Over time, this gasket will require replacement.

Check the shroud gasket (AZI P/N 300-0446) regularly for cracks or deterioration.

Perform a Kimwipe test to verify the seal’s integrity. (See Kimwipe test below.)

If Kimwipe pulls out easily, then either the shroud gasket or insulator seal will need replacement. (See Insulator Seal, Section 10.3 below, for more information)

If replacement is needed, contact AZI customer service at 800-528-7411, 602-470-1414, or by e-mail at support@azic.com for further details and information.

Kimwipe Test:

Place a Kimwipe, single thickness, on the ceramic insulator plate at the 3 o’clock position.

Close lid.

Attempt to pull Kimwipe out from between the gasket and plate, being careful not to rip the Kimwipe.

Repeat at the 6 o’clock and 9 o’clock positions.

oIf it removes easily, then either the shroud gasket or insulator seal--or

both--need to be replaced.

oIf it does not pull out, then the shroud gasket and insulator seal are o.k.

Manual backgroundManual background 10.3. Insulator Seal

Beneath the ceramic insulator plate is an insulator seal that should be checked for compression. The above Kimwipe test should be performed. If the Kimwipe is removed easily, then the insulator seal needs to be replaced. To replace:

Switch the main power switch OFF.

Remove the pan support stem and ceramic insulator plate.

Replace the worn seal (AZI P/N 250-0001).

Reassemble.

AZI Customer Service 800-528-7411 or 602-470-1414

Page 79 of 92