Arizona MAX-5000XL user manual Moistures too high or

Models: MAX-5000XL

1 92
Download 92 pages 19.89 Kb
Page 84
Image 84

Moistures too high or

Temperature set too high

Lower temperature,

solids too low

 

see pages 25 & 68

 

 

 

 

 

 

Bent sample pan

Use new pan

 

 

 

 

Dirty sample pan

Use new pan

 

 

 

 

Debris in balance gaps

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

 

Dirty heater assembly

Clean heater assembly,

 

 

see page 80

 

 

 

 

Ending criteria incorrect

Determine new ending

 

 

criteria, see page 26

 

 

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

Moistures too low or

Temperature too low

Increase temperature,

solids too high

 

see pages 25 & 68

 

 

 

 

 

 

Sample size too small or too

Adjust sample size,

 

large

see pages 29 & 66

 

 

 

 

 

 

Sample needs spreading

Spread sample evenly,

 

 

see page 65

 

 

 

 

Ending criteria incorrect

Determine new ending

 

 

criteria, see page 26

 

 

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

Moisture never stops

Other volatiles

Try lower temperature,

increasing

 

see pages 29 & 66

 

 

 

 

 

 

Balance drifting

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

No graphical display

Graphics not selected

Press [GRAPH] key at Main

 

 

Test Screen

 

 

 

No heat

Voltage selection wrong

Check power module at rear

 

 

of instrument, set to correct

 

 

voltage. See page 14

 

 

 

 

Wiring failure in heater circuit

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

 

Heater failure

Call AZI at 800-528-7411 or

 

 

602-470-1414 for assistance

 

 

 

AZI Customer Service 800-528-7411 or 602-470-1414

Page 84 of 92

Page 84
Image 84
Arizona MAX-5000XL user manual Moistures too high or