5.3. Calibration Menu
Balance Calibration
Balance calibration is explained and performed as part of instrument setup, see page
20.The sequence is repeated here for continuity. The balance should be calibrated:
∙If the instrument has been turned off for several hours or in storage,
∙After the instrument is moved,
∙Whenever a problem is suspected, and/or
∙As often as required by your quality policy.
NOTE: The MAX 5000XL must be powered on until the balance temperature stability icon in the lower left corner of the main screen changes from unstable to stable to allow the balance to stabilize before beginning the balance calibration. (See section 6.1 Balance Temperature Stability on page 64 for more information). The unit should not be moved, jarred, or otherwise disturbed in any way during the calibration period. A
Calibrate Balance
∙From the Main Test Screen, press [Menu].
∙Select the CALIBRATION MENU.
∙Select BALANCE CALIBRATION.
∙Select CALIBRATE BALANCE.
∙Press [GO]. Follow the prompts to complete the calibration.
∙The first prompt is “OPEN LID” and “PLACE CLEAN PAN ON PAN SUPPORT”
∙After placing a clean pan on the pan support and closing the LID, the screen will
display “CALIBRATING, PLEASE WAIT.”
∙As soon as the tare is finished, the next prompt, “OPEN LID” and “LOAD 50G WEIGHT AND WAIT,” will appear.
∙As the lid is opened, the screen prompts: “LOAD 50G WEIGHT AND WAIT.”
Avoid excess pressure on the scale. Use the tweezers included with the MAX 5000XL to gently place the weight on the pan.