Arizona MAX-5000XL user manual Report Setup, Report Control Options

Models: MAX-5000XL

1 92
Download 92 pages 19.89 Kb
Page 51
Image 51
 Report Setup

Report Control OptionsManual background Report Setup

This menu allows the report outputs to be enabled and the printout to be customized to your needs.

Report Control Options

REPORT ENABLED – A YES sends a progress report (optional, see below) to the printer during a test and a final report to

the printer after a test.

REPORT START – These three options determine if a progress report begins at the start of the test, when the first prediction of final result is made, or just a final report at the end of the test.

REPORT INTERVAL – This selection determines how often the progress report data is sent to the printer or computer. Times of 1, 5, 10, 15, and 30 seconds or one (1) minute are available. The short durations are good for trials and research and the long intervals are good for normal use.

INCLUDE GRAPH – When YES is set, a graph of the results will be sent to the printer after the final report. Graphic data is not sent to the computer.

AZI Customer Service 800-528-7411 or 602-470-1414