Reference Manual

Section 2: Configuration

00809-0100-4102, Rev AA

May 2013

2.3.1Connection diagrams

Bench hook-up

Connect the bench equipment as shown in Figure 2-1 on page 11, and turn on the Field Communicator by pressing the ON/OFF key or log into AMS. The Field Communicator or AMS will search for a HART-compatible device and indicate when the connection is made. If the Field Communicator or AMS fail to connect, it indicates that no device was found. If this occurs, refer to Section 6: Troubleshooting.

Field hook-up

Figure 2-1 on page 11 illustrates the wiring for a field hook-up with a Field Communicator or AMS. The Field Communicator or AMS may be connected at “COMM” on the transmitter Power Module.

Figure 2-1. Field Communicator Connection

For HART communication, a 2051 WirelessHART DD is required.

2.4Basic setup

2.4.1Set device tag

Fast Keys

2, 2, 9, 1, 1

The tag is used to identify the device. You can use an 8 to 32 character tag.

 

1.

From the Home screen, select 2: Configure

 

 

2.

Select 2: Manual Setup

 

 

3.

Select 9: Device Information

 

 

4.

Select 1: Identification

 

Configuration

5.

Select 1: Tag

11

 

 