Reference Manual

Appendix A: Reference Data

00809-0100-4102, Rev AA

May 2013

 

 

Rosemount 2051L

Limit is flange rating or sensor rating, whichever is lower (see the table below).

Standard

 

Type

CS Rating

 

SST Rating

 

 

 

 

 

 

ANSI/ASME

 

Class 150

285 psig

 

275 psig

 

 

 

 

 

 

ANSI/ASME

 

Class 300

740 psig

 

720 psig

 

 

 

 

 

 

ANSI/ASME

 

Class 600

1480 psig

 

1440 psig

 

 

 

 

 

 

 

 

At 100 °F (38 °C), the rating decreases

 

 

with increasing temperature, per ANSI/ASME B16.5.

 

 

 

 

 

 

 

DIN

 

PN 10–40

40 bar

 

40 bar

 

 

 

 

 

 

DIN

 

PN 10/16

16 bar

 

16 bar

 

 

 

 

 

 

DIN

 

PN 25/40

40 bar

 

40 bar

 

 

 

 

 

 

At 248 °F (120 °C), the rating decreases with increasing temperature, per DIN 2401.

A.3.2 Static pressure limit

Rosemount 2051CD Only

Operates within specifications between static line pressures of 0.5 psia and 3626 psig (4500 psig (310, 3 bar) for

Option Code P9).

Range 0: 0.5 psia and 750 psig (3, 4 bar and 51, 7 bar)

Range 1: 0.5 psia and 2000 psig (3, 4 bar and 137, 9 bar)

A.3.3 Burst pressure limits

Coplanar or traditional process flange

10000 psig (689,5 bar).

Rosemount 2051T

Ranges 1–4: 11000 psi (758,4 bar)

Range 5: 26000 psig (1792,64 bar)

A.3.4 Temperature limits

Ambient

–40 to 185 °F (–40 to 85 °C)

With LCD display(1): –40 to 175 °F (–40 to 80 °C)

(1) LCD display may not be readable and LCD updates will be slower at temperatures below -4 °F (-20 °C).

Storage

–40 to 185 °F (–40 to 85 °C)

With LCD display: –40 to 185 °F (–40 to 85 °C)

Specifications and Reference Data

95

Page 107
Image 107
Emerson 00809-0100-4102 manual Static pressure limit, Burst pressure limits, Temperature limits