Section 5: Operation and Maintenance

Reference Manual

May 2013

00809-0100-4102, Rev AA

5.4.3Recall Factory Trim—Sensor Trim

The Recall Factory Trim—Sensor Trim command allows the restoration of the as-shipped factory settings of the Sensor Trim. This command can be useful for recovering from an inadvertent zero trim of an absolute pressure unit or inaccurate pressure source.

Recalling factory trim with AMS

Right click on the device and, under the Method drop down menu, move cursor over Calibrate and select Restore Factory Calibration.

1.Click Next after setting the control loop to manual.

2.Select Sensor Trim under Trim to recall and click Next.

3.Follow the screen prompts to recall Sensor Trim.

5.4.4Line Pressure Effect (Range 2 and Range 3)

The following specifications show the static pressure effect for the Rosemount 2051 Range 2 and Range 3 pressure transmitters used in differential pressure applications where line pressure exceeds 2000 psi (138 bar).

Zero Effect

±0.1% of the upper range limit plus an additional ± 0.1% of upper range limit error for each 1000 psi (69 bar) of line pressure above 2000 psi (138 bar).

Example: Line pressure is 3000 psi (207 bar) for Ultra performance transmitter. Zero effect error calculation:

± {0.05 + 0.1 x [3 kpsi - 2 kpsi]} = ± 0.15% of the upper range limit

Span Effect

Refer to “Line Pressure Effect” on page 91.

5.4.5Compensating for Line Pressure (Range 4 and Range 5)

The Rosemount 2051 Wireless Range 4 and 5 pressure transmitters require a special calibration procedure when used in differential pressure applications. The purpose of this procedure is to optimize transmitter performance by reducing the effect of static line pressure in these applications. The 2051 Wireless differential pressure transmitters (Ranges 1, 2, and 3) do not require this procedure because optimization occurs in the sensor.

Applying high static pressure to the 2051 Wireless Range 4 and Range 5 pressure transmitters causes a systematic shift in the output. This shift is linear with static pressure; correct it by performing the “Sensor Trim” procedure on page 68.

The following specifications show the static pressure effect for the 2051 Wireless Range 4 and Range 5 transmitters used in differential pressure applications:

70

Operation and maintenance

Page 82
Image 82
Emerson 00809-0100-4102 manual Recall Factory Trim-Sensor Trim, Line Pressure Effect Range 2 and Range