Section 3: Installation

Reference Manual

May 2013

00809-0100-4102, Rev AA

Bolt installation

Only use bolts supplied with the Rosemount 2051 or sold by Emerson Process Management as spare parts. When installing the transmitter to one of the optional mounting brackets, torque the bolts to 125 in-lb. (0,9 N-m). Use the following bolt installation procedure:

1.Finger-tighten the bolts.

2.Torque the bolts to the initial torque value using a crossing pattern.

3.Torque the bolts to the final torque value using the same crossing pattern. Torque values for the flange and manifold adapter bolts are as follows:

Table 3-2. Bolt Installation Torque Values

 

Bolt Material

 

Initial Torque Value

Final Torque Value

 

 

CS-ASTM-A445 Standard

 

300 in.-lb (34 N-m)

650 in.-lb (73 N-m)

 

 

316 SST—Option L4

 

150 in.-lb (17 N-m)

300 in.-lb (34 N-m)

 

 

ASTM-A-193-B7M—Option L5

 

300 in.-lb (34 N-m)

650 in.-lb (73 N-m)

 

 

Alloy K-500—Option L6

 

300 in.-lb (34 N-m)

650 in.-lb (73 N-m)

 

 

ASTM-A-453-660—Option L7

 

150 in.-lb (17 N-m)

300 in.-lb (34 N-m)

 

 

ASTM-A-193-B8M—Option L8

 

150 in.-lb (17 N-m)

300 in.-lb (34 N-m)

 

 

 

 

 

 

 

Figure 3-6. Traditional flange bolt configurations

 

 

 

A

 

B

C C

D C

Note

E

F

E

F

Dimensions are in inches

 

A. Differential Transmitter

B. Gage/Absolute Transmitter

C.Drain/Vent

D.Vented fitting

E.1.75 (44) × 4

F.1.50 (38) × 4(1)

(1)For Gage and Absolute Transmitters: 150 (38) x 2