Section 5: Operation and Maintenance

Reference Manual

May 2013

00809-0100-4102, Rev AA

 

 

c o n f g

w a r n

Diagnostic Button Screen 9.2: configuration of the transmitter is invalid such that non-critical operation of the device may be affected - check the extended configuration status to identify which configuration item(s) need to be corrected

s n s r

f a i l u r

Diagnostic Button Screen 10.1: a sensor attached to the transmitter has failed, and valid readings from that sensor are no longer possible - check the sensor and sensor wiring connections - check additional status for more detailed information of the failure source

s n s r

w a r n

Diagnostic Button Screen 10.2: a sensor attached to the transmitter is degraded, readings from that sensor may not be within accuracy specifications - check the process, and sensor wiring connections - check additional status for more detailed information of the warning source

Note

Use the Rosemount Wireless LCD Part Number: 00753-9004-0002.

82

Operation and maintenance

Page 94
Image 94
Emerson 00809-0100-4102 manual N f g R n