HYDRA

Service Manual

The output of the buffer is ac-coupled by A3C15 and A3C16 to the true-rms ac-to-dc converter A3U6. Discharge JFET A3Q13 is switched on to remove any excess charge from the coupling capacitors A3C15 and A3C16 between channel measurements. A3C17 provides an averaging function for the converter, and resistor network A3Z1 divides the output by 2.5 before sending the signal to the active ac volts filter. Analog processor switch S81 connects the output of the active filter to HI of the A/D Converter. Components A3R29, A3R30, A3C26, and A3C28 provide filtered power supplies (+VAC and -VAC) for the ac buffer, the ac switch JFETs, and the rms converter.

2A-56.Frequency

After any dc component is blocked by capacitors A3C15, A3C16, and A3C31, the output of the ac buffer is used to determine the input frequency. This signal is sent to the ACBO pin of analog processor A3U8 and switched to the internal frequency comparator and counter circuit by S42.

2A-57.Passive and Active Filters

The passive filters are used for the dc voltage and ohms measurements. For most ranges, capacitors A3C14 and A3C11 are switched into the measurement circuit in front of the A3U8 A/D Converter by switches S86, S87, and S88. These capacitors act with the 100- kΩ series resistance provided by A3R42 or A3Z4 to filter out high-frequency noise. For the 300-kΩrange, only A3C14 is switched in by switches S86 and S85. For the 3-MΩand 10-MΩranges, A3C11 or A3C14 are not switched in to keep settling times reasonably short.

Between channel measurements, the passive filters are discharged by JFET A3Q2 under control of Microcontroller A3U9 through comparator A3U14. When the ZERO signal is asserted, A3R14 pulls the gate of A3Q2 to ground, turning the JFET on and discharging A3C11. At the same time, zeroing of filter capacitors A3C14 and A3C27 is accomplished by having the Analog Processor turn on internal switches S2, S86, and S87.

The active filter is only used for ac voltage measurements. This three-pole active filter removes a significant portion of the ac ripple and noise present in the output of the rms converter without introducing any additional dc errors. The active filter op-amp within A3U8, resistors A3R20, A3R17, and A3R16, and capacitors A3C7, A3C10, and A3C6 form the filter circuit. This filter is referenced to the LO input to the a/d converter within A3U8 by the op-amp. The input to the filter is available at the RMSO pin, and the output is sent to the RMSF pin of A3U8. Switches S80 and S82, which are turned on prior to each new channel measurement, cause the filter to quickly settle (pre-charge) to near the proper dc output level.

2A-58.A/D Converter

Figure 2A-7 shows the dual slope a/d converter used in the instrument. The unknown input voltage is buffered and used to charge (integrate) a capacitor for an exact period of time. This integrator capacitor is then discharged by the buffered output of a stable and accurate reference voltage of opposite polarity. The capacitor discharge time, which is proportional to the level of the unknown input signal, is measured by the digital circuits in the Analog Measurement Processor. This time count becomes the conversion result.

2A-30

Page 112
Image 112
Fluke 2635A, 2625A service manual 2A-57.Passive and Active Filters, 2A-58.A/D Converter

2635A, 2625A specifications

The Fluke 2625A is a sophisticated data logger designed for a wide range of industrial and laboratory applications. This versatile instrument excels in capturing and monitoring temperature and voltage measurements with precision and reliability. One of its standout features is its ability to record data from various sensors, making it ideal for environments where comprehensive data acquisition is critical.

Equipped with a large LCD display, the Fluke 2625A provides clear and user-friendly access to real-time measurement data. Its graphical interface allows users to visualize trends and monitor readings easily, which significantly enhances usability. The device supports a variety of input types and can connect to thermocouples, RTDs, and other sensors, offering great flexibility for various measurement tasks.

The Fluke 2625A employs advanced measurement technologies that ensure accuracy in data logging. With a high sampling rate, it captures fast-changing signals while maintaining data integrity. The device features a built-in memory that allows for extensive logging, accommodating long-term monitoring tasks without requiring constant oversight. Users can set it to log data at specified intervals, offering customizable solutions for different monitoring needs.

Another significant characteristic of the Fluke 2625A is its robust communication capabilities. The device supports USB and RS-232 interfaces for easy connectivity with computers and other devices, making data transfer and analysis seamless. This feature is especially beneficial for users who need to compile or analyze large sets of data quickly.

Furthermore, the Fluke 2625A is designed with durability in mind. Its rugged construction makes it suitable for challenging environments, including those with high levels of vibration, dust, or moisture. This reliability ensures that the device can withstand the rigors of fieldwork while still delivering accurate measurements.

In summary, the Fluke 2625A is a powerful data logger that combines versatility, accuracy, and reliability. Its user-friendly interface, advanced measurement technologies, and robust design make it an essential tool for professionals in various industries, from manufacturing and HVAC to research and development. Whether monitoring temperature changes in a laboratory or assessing voltage levels in an industrial setting, the Fluke 2625A stands out as a trusted solution for effective data logging and analysis.